Functions
7.10 Float transmission
FC410 with Modbus
78
Operating Instructions, 02/2016, A5E33120874-AC
This table shows the different options for setting the transmission method:
Selection
Sequence
1st
2nd
3rd
4th
1 - 0 - 3 - 2
Byte 1
(MMMMMMMM)
Byte 0
(MMMMMMMM)
Byte 3
(SEEEEEEE)
Byte 2
(EMMMMMMM)
0 - 1 - 2 - 3
Byte 0
(MMMMMMMM)
Byte 1
(MMMMMMMM)
Byte 2
(EMMMMMMM)
Byte 3
(SEEEEEEE)
2 - 3 - 0 - 1
Byte 2
(EMMMMMMM)
Byte 3
(SEEEEEEE)
Byte 0
(MMMMMMMM)
Byte 1
(MMMMMMMM)
3 - 2 - 1 - 0 *
Byte 3
(SEEEEEEE)
Byte 2
(EMMMMMMM)
Byte 1
(MMMMMMMM)
Byte 0
(MMMMMMMM)
* = Factory setting
S = Sign
E = Exponent
M = Mantissa
NOTICE
Change of float byte order
If the float byte order is changed by use of PDM to anything other than default, all float
values shown in PDM are wrong.
Содержание SITRANS FC410
Страница 2: ......
Страница 70: ...Commissioning 6 14 Process variables FC410 with Modbus 68 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 92: ...Troubleshooting FAQs 10 2 Troubleshooting FC410 with Modbus 90 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 142: ...Sensor dimension dependent default settings FC410 with Modbus 140 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 146: ...Zero point adjustment FC410 with Modbus 144 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 150: ...CRC calculation FC410 with Modbus 148 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 152: ...Exception codes E 1 Exception handling FC410 with Modbus 150 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 154: ...Float definition F 1 Float definition FC410 with Modbus 152 Operating Instructions 02 2016 A5E33120874 AC ...