Modbus holding registers
A.3 Modbus holding registers tables
FC410 with Modbus
134
Operating Instructions, 02/2016, A5E33120874-AC
Table A- 16 Massflow calibration
Modbus
address
Data type /
Size [bytes]
Parameter
Description
Default value
[unit]
Value range Access
level
2101
Float / 4
Maximum Massflow
Capacity
Maximum massflow
measurement capacity of
the sensor
Sensor size
specific [kg/s]
1)
0 to 1023
Read only
2402
Float / 4
Calibration Factor
Factory-set sensor-specific
calibration factor.
The calibration factor is
shown on the sensor name-
plate
-
Min:
5.00E+07
Max:
4.29E+09
Read only
1)
: See Sensor dimension dependent default settings (Page 139).
Table A- 17 Density calibration
Modbus
address
Data type /
Size [bytes]
Parameter
Description
Default value
[unit]
Value range Access
level
2428
Float / 4
Density Calibration
Offset
Specify an offset in the
density flow calculation
-
-14 000 to
+14 000
Read only
2430
Float / 4
Density Calibration
Factor
Specify gain factor in the
density flow calculation
-
-1-999 to
+1.999
Read only
2432
Float / 4
Dens. Comp. Tube
Temp.
Specifies a tube tempera-
ture coefficient in the densi-
ty calculation
-
-0.001953
to
+0.001953
Read only
2434
Float / 4
Dens. Comp. Frame
Temp.
Specifies a frame tempera-
ture coefficient in the densi-
ty calculation
-
-0.001953
to
+0.001953
Read only
Содержание SITRANS FC410
Страница 2: ......
Страница 70: ...Commissioning 6 14 Process variables FC410 with Modbus 68 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 92: ...Troubleshooting FAQs 10 2 Troubleshooting FC410 with Modbus 90 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 142: ...Sensor dimension dependent default settings FC410 with Modbus 140 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 146: ...Zero point adjustment FC410 with Modbus 144 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 150: ...CRC calculation FC410 with Modbus 148 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 152: ...Exception codes E 1 Exception handling FC410 with Modbus 150 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 154: ...Float definition F 1 Float definition FC410 with Modbus 152 Operating Instructions 02 2016 A5E33120874 AC ...