POLYMOBIL Plus
SPR8-125.061.02
Page 4 of 4
Siemens AG
Rev. 01
01.00
TD SD 24
Medical Engineering
1 - 4
Prerequisites
•
Connect the POLYMOBIL Plus only to a line voltage supply (line voltage receptacle) that
complies with the requirements of VDE 0107, or corresponds to the local national
regulations.
•
Disconnect the POLYMOBIL Plus at the line voltage
OFF
switch on the operating
console and disconnect the
line voltage plug
prior to any service work
.
•
Remove or insert boards with the generator switched OFF only; observe
ESD guidelines when handling boards.
Replacing damaged or missing screws
1
•
Damaged or missing screws may only be replaced by steel screws as specified in the
installation drawings conforming to DIN267 and must have the specified tensile strength.
Cleaning
1
•
The unit must always be switched off or disconnected before cleaning.
•
Use only water to clean the unit or a lukewarm, dilute solution consisting of water and a
household cleaning agent.
•
Do not use abrasive cleaning agents or organic solvents such as benzine, alcohol or spot
remover. Do not spray with water.
•
For additional information, refer to the Operating Instructions "Cleaning/Disinfection".
All Allen screws must have a tensile strength rating of 8.8.
CAUTION