POLYMOBIL Plus
SPR8-125.061.02
Page 12 of 20
Siemens AG
Rev. 01
01.00
TD SD 24
Medical Engineering
5 - 12
Measurements
Setting the maximum filament inverter frequency
5
The maximum filament inverter freqency is 25 kHz and can be set with P3 on D915.
Proceed as follows:
•
Switch POLYMOBIL Plus
OFF.
•
Remove fuse F6 on D925.
•
Connect the oscilloscope to D915 at TP "CAL" and "GND".
•
Switch POLYMOBIL Plus
ON
(Err 03 is blinking)
.
The oscillogram must appear like the one in Fig. 12.
•
Setting the maximum filament inverter frequency with potentiometer P3 on D915.
Fig. 12
CH1: D915 test point "CAL"
20 µs /Div