Troubleshooting
3
Error messages
3
a) During initialization
b)During stand by
c)During exposure
Error messages Err 1, and initialization errors Err 90, Err 96 and
Err 97 block the generator so that the system must be switched
off and switched back on again.
"StandBy" error messages disappear when the error is cleared.
Error messages which occur during an exposure will continue
flashing until another key is activated.
Err 90
EPROM Checksum
Err 96
KVnom error
Err 97
mAnom error
Err 1
Capacity charge error
Charging circuit error
Err 2
+15 VError
Err 3
Vfil < Vstby
Err 4
Vfil > Vstby
Err 5
KVact
≠
0
Err 6
mAact
≠
0
Err 7
Rot
≠
0
Err 8
Braking failure
Error in the rotating anode circuit
Err 9
Pressure switch
Oil pressure switch in the single tank
is active
Err 30
Fast capacitor charge
Capacitor bank is charging too fast
Err 31
Slow capacitor charge
Capacitor bank is charging too slow
Err 33
Inverter Error
Err 34
Filament circuit Inverter Short-Cir-
cuit
Err 10
Rot
=
0
No rotating anode acceleration
Err 11
Short circuit in main inverter
Short circuit in the main inverter
Err 12
KVmax
Err 13
Imax
Err 14
KVact < KVnom
Err 15
mAact < mAnom
Err 17
Maximum exposure time
Max. exposure reached
Err 18
Exposure not completed
Exposure time not completed(expo-
sure release switch released)
Err 20
Capacitor not ready
Capacitor bank not completely
charged
Err 21
Vfil > Vmax during prep
Maximum filament heating exceeded
during preparation
Err 22
Maximum prep time
Max. preparation time exceeded