POLYMOBIL Plus
SPR8-125.061.02
Page 2 of 2
Siemens AG
Rev. 01
01.00
TD SD 24
Medical Engineering
4 - 2
Service programs
Service programs available
4
Pr. 1
Configures the capacitor bank.
After starting the service program,"CAP 150" appears on the display.
The voltage of the capacitor increases approximatelly in increments from
150 V to 250 V, 330 V, 400 V, 430 V up to 440 V.
The program remains at each step for about 6 minutes at the 150 V to
440 V level.
The system remains at the 440 V level for one hour, after which
the program automatically ends; "CAP END" appears on the display.
Pr. 2
Reads out the exposure counter.
Pr. 3
Error memory with up to 20 errors.The left display indicates how many
errors there are.The error message is displayed in the right display. The-
latest error is displayed first.
Pr. 4
Deletes error memory.
Press mAs + key for approximately 4 seconds.
Pr. 5
Sets the default exposure parameters (according to customer request).
Pr. 6
Sets the maximum possible KV and mAs values(according to customer
request.
Pr. 7
Adjusts FMAX on the main inverter.
"ADJ F" appears on the display after starting the service program.
Press S27.
"F on" appears on the display.
FMAX can be measured at TP "REG" on D 915, and can be set with the
potentiometer P4.
Pr. 8
Tests the anode starter unit.
Press S 27.
Pr. 9
Tests the filament unit.
"FIL 3" appears on the display after starting the service program.
Press S27.
Filament boosts; "FIL 5" appears on the display.
There is no radiation present during these service programs.