7ML19981FB05
MultiRanger 100/200 – INSTRUCTION MANUAL
Page 227
m
mmmm
Ap
pendix B:
T
e
c
h
R
e
fe
re
nce
Scanning
MultiRanger 100/200
When echo processing is complete (if more than 1 vessel is monitored) the scanning relay
changes state to supply the transmit pulse to the other transducer after the Scan Delay
(P727).
Scan Delay is automatically set by Maximum Process Speed (P003). When high speed
scanning is required (sometimes the case for equipment position monitoring), the Scan
Delay may be reduced. Reduce the Scan Delay only as required, otherwise premature
scanning relay fatigue could occur.
MultiRanger 200
When two transducers are connected and configured in a dual point unit, the
MultiRanger will scan each in turn via the scanner relay. When a single point
MultiRanger is programmed for differential or average level Operation (P001 = 4 or 5), two
transducers of the same type must be used.
Volume Calculation [MR 200]
Please note that the Volume Calculation feature is only available on the MultiRanger 200.
The unit provides a variety of volume calculation features (P050 to P055).
If the vessel does not match any of the eight preset Tank Shape calculations, a Universal
Volume calculation may be used. Use the level/volume graph or chart provided by the
vessel fabricator (or create one based on the vessel dimensions). Based on the graph,
choose the Universal Volume calculation, and select the level vs. volume breakpoints to
be entered (32 max). Generally, the more breakpoints entered, the greater the accuracy.
Universal, Linear (P050 = 9)
This volume calculation creates a piece-wise linear approximation of the level/volume
curve. This option provides best results if the curve has sharp angles joining relatively
linear sections.
Enter a Level Breakpoint at each point where the level/volume curve bends sharply (2
minimum).
For combination curves (mostly linear but include one or more arcs), enter numerous
breakpoints along the arc, for best volume calculation accuracy.
Содержание MAG 1100
Страница 1: ... MFDUSPNBHOFUJD GMPXNFUFST 1 8 0QFSBUJOH OTUSVDUJPOT p 4 53 4 ...
Страница 47: ......
Страница 76: ... MFDUSPNBHOFUJD GMPXNFUFST 1 8 0QFSBUJOH OTUSVDUJPOT p 4 53 4 ...
Страница 135: ...60 SITRANS F M 7 Electrical connection 19 IP 20 version EEx e 19 IP 20 version 19 IP 66 version ...
Страница 140: ...65 SITRANS F M 8 Commissioning 8 3 2 MAG 5000 CT and MAG 6000 CT Menu overview ...
Страница 164: ...Tab 36 Siemens MultiRanger 100 ...
Страница 165: ...Instruction Manual April 2006 100 200 multiranger ...
Страница 258: ...Page 84 MultiRanger 100 200 INSTRUCTION MANUAL 7ML19981FB05 mmmm m Open Channel Monitoring MR 200 ...
Страница 396: ...Page 222 MultiRanger 100 200 INSTRUCTION MANUAL 7ML19981FB05 mmmmm Parameters ...
Страница 445: ...Operation Manual February 2005 ECHOMAX xrs 5 ...
Страница 446: ...IQ300IX fm Page 5 Tuesday October 2 2001 1 43 PM ...
Страница 454: ...Page EN 8 XRS 5 Series Transducers OPERATION MANUAL 7ML19985QT81 mmmmm English ...
Страница 462: ...Side DA 8 XRS 5 Serie Transducere DRIFTSVEJLEDNING 7ML19985QT81 mmmmm Dansk ...
Страница 470: ...Seite 8 Ultraschallsensoren XRS 5 BETRIEBSANLEITUNG 7ML19981QT81 mmmmm Deutsch ...
Страница 478: ...Σελίδα EL 8 Μορφοτροπείς σειράς XRS 5 ΕΓΧΕΙΡΙ ΙΟ ΛΕΙΤΟΥΡΓΙΑΣ 7ML19985QT81 mmmmm Ελληνικά ...
Страница 486: ...Página ES 8 Sensores XRS 5 MANUAL DE OPERACIÓN 7ML19985QT81 mmmmm Español ...
Страница 494: ...Page FR 8 Transducteurs Série XRS 5 MANUEL D UTILISATION 7ML19985QT81 mmmmm Français ...
Страница 502: ...Pagina IT 8 Trasduttori serie XRS 5 MANUALE OPERATIVO 7ML19985QT81 mmmmm Italiano ...
Страница 510: ...Pagina NE 8 XRS 5 serie transducers BEDIENINGSHANDLEIDING 7ML19985QT81 mmmmm Nederlands ...
Страница 518: ...Página PT 8 Transdutores Série XRS 5 MANUAL DE FUNCIONAMENTO 7ML19985QT81 mmmmm Português ...
Страница 526: ...sivu SU 8 XRS 5 sarjan anturit KÄYTTÖOPAS 7ML19985QT81 mmmmm Suomi ...
Страница 534: ...Sida SV 8 XRS 5 Transducerserier ANVÄNDARMANUAL 7ML19985QT81 mmmmm Svenska ...
Страница 539: ...Operation Manual February 2005 ECHOMAX xps xct ...
Страница 540: ...IQ300IX fm Page 5 Tuesday October 2 2001 1 43 PM ...
Страница 550: ...Page EN 10 XPS XCT Series Transducers OPERATION MANUAL 7ML19985QM82 English ...
Страница 555: ...Tab 37 Siemens XRS 5 Ultrasonic Level Sensors ...
Страница 556: ...Operation Manual February 2005 ECHOMAX xrs 5 ...
Страница 557: ...IQ300IX fm Page 5 Tuesday October 2 2001 1 43 PM ...
Страница 565: ...Page EN 8 XRS 5 Series Transducers OPERATION MANUAL 7ML19985QT81 mmmmm English ...
Страница 573: ...Side DA 8 XRS 5 Serie Transducere DRIFTSVEJLEDNING 7ML19985QT81 mmmmm Dansk ...
Страница 581: ...Seite 8 Ultraschallsensoren XRS 5 BETRIEBSANLEITUNG 7ML19981QT81 mmmmm Deutsch ...
Страница 589: ...Σελίδα EL 8 Μορφοτροπείς σειράς XRS 5 ΕΓΧΕΙΡΙ ΙΟ ΛΕΙΤΟΥΡΓΙΑΣ 7ML19985QT81 mmmmm Ελληνικά ...
Страница 597: ...Página ES 8 Sensores XRS 5 MANUAL DE OPERACIÓN 7ML19985QT81 mmmmm Español ...
Страница 605: ...Page FR 8 Transducteurs Série XRS 5 MANUEL D UTILISATION 7ML19985QT81 mmmmm Français ...
Страница 613: ...Pagina IT 8 Trasduttori serie XRS 5 MANUALE OPERATIVO 7ML19985QT81 mmmmm Italiano ...
Страница 621: ...Pagina NE 8 XRS 5 serie transducers BEDIENINGSHANDLEIDING 7ML19985QT81 mmmmm Nederlands ...
Страница 629: ...Página PT 8 Transdutores Série XRS 5 MANUAL DE FUNCIONAMENTO 7ML19985QT81 mmmmm Português ...
Страница 637: ...sivu SU 8 XRS 5 sarjan anturit KÄYTTÖOPAS 7ML19985QT81 mmmmm Suomi ...
Страница 645: ...Sida SV 8 XRS 5 Transducerserier ANVÄNDARMANUAL 7ML19985QT81 mmmmm Svenska ...
Страница 650: ...Tab 38 Siemens XPS XCT Transducers ...
Страница 651: ...Operation Manual February 2005 ECHOMAX xps xct ...
Страница 652: ...IQ300IX fm Page 5 Tuesday October 2 2001 1 43 PM ...
Страница 662: ...Page EN 10 XPS XCT Series Transducers OPERATION MANUAL 7ML19985QM82 English ...