LITHOSTAR Multiline
RXL2-120.812.01
Page 24 of 24
Siemens AG
System
Rev. 02
10.04
CS PS 24
Medical Solutions
4 - 24
Connections with POLYDOROS SX 65
Fastening of plug, Sonoline
4
•
Remove the existing screw at point (1/Fig. 17).
•
Screw in the mounting screw according to Fig. 17. The clamp (2/Fig. 17) prevents
loosening the plug connection without tools.
•
Tighten the counternut (1/Fig. 17).
Fig. 17
CE approval is valid for the entire system only. If the ultrasound
unit is removed, then this does not retain the required CE appro-
val.
1
2
NOTE