Connections with POLYDOROS SX 65
4 - 19
Siemens AG
RXL2-120.812.01
Page 19 of 24
LITHOSTAR Multiline
Medical Solutions
Rev. 02
10.04
CS PS 24
System
Connections/harness locations in the generator cabinet (Part C)
4
•
Route the harnesses in the generator cabinet as shown in Fig. 13 above, and connect
them in accordance with Table 10:
Generator power line connection
Nominal Line Voltage: 400 V ± 10%
•
Secure the screw type terminals from the service packet to the conductors of the power
cable. If the power cable diameter is small enough, turn the clamps 180° (both
curvatures opposite to each other).
•
Apply end splices to the power cable.
•
Connect power leads L1, L2, L3, N, PE (connection
/Fig. 13) to terminal strip K20.
•
Strain-relieve the power cable. Existing shielding should not be connected.
Generator - Periphery
Type of Cable
Function
Generator
ok
Power line cable
3/N/Gnd
System
power line connection
M16. K20
Main distributor
L1, L2, L3, N,
2 x 0.75 mm
2
Radiation protection
9)
M16.D160.X61 Door
contact
2 x 0.75 mm
2
Radiation buzzer
Radiation display
9)
M16.D160.X61 X-ray
display
lamps
Tab. 10
9)
option
J
J
J
A
A