Connections with POLYDOROS SX 65
4 - 17
Siemens AG
RXL2-120.812.01
Page 17 of 24
LITHOSTAR Multiline
Medical Solutions
Rev. 02
10.04
CS PS 24
System
•
Connect the protective conductor cable to H1.
(2/Fig. 12).
•
Thread the (primary) leads H1.U1 and H1.V1 of the inverter module per the designation
onto H1 (3/Fig. 13).
•
Route the cable connected to D220.X41 (4/Fig. 13) to board D160 and connect it to plug
X44.
•
Plug the cable X1 from board D100 to connector X1 on board D220 (5/Fig. 12) and
strain-relieve it on the cover (shielding underneath the strain relief (6/Fig. 12)).
•
Reinstall the guard with the
symbols (1/Fig. 12).
•
Check the nuts on the protective conductor studs (7/Fig. 13) and tighten them, if neces-
sary.