Connections with POLYDOROS SX 65
4 - 15
Siemens AG
RXL2-120.812.01
Page 15 of 24
LITHOSTAR Multiline
Medical Solutions
Rev. 02
10.04
CS PS 24
System
Connecting the H1 High Voltage Transformer
•
Remove the inverter cover and open the bayonet connectors (1/Fig. 10).
•
Change the position of the transformer mount (Fig. 10) and swing it out.
•
Place the high voltage transformer on the transformer mount.
•
Route the high voltage harnesses up to the high voltage transformer (Pfeil/Fig. 11).
•
Check the oil level in the high voltage receptacles (approximately 1 cm); if necessary,
refill with oil (part no. 44 10 023 RV090 1 liter).
•
Insert the high voltage plug without silicone disk but with silicone gasket into the correct
receptacle of H1 (Workstation 1).
•
Slide the H1 high voltage transformer into the generator cabinet.
Generator - Unit Connections (Yellow Cable Harness )
Type of Cable
Function
Generator
System
ok
Stator cable
3 x 0.75 mm
2
Rotating anode drive
Anode starter
K31.R2, R4, K32.R2
X-ray tube I, 0, II
2 x HSK
HV power supply
HV-Transf. H1.API
X-ray tube
1 x 6 mm
2
Protective conductor
- bolts
(on M16 under K20)
X-ray tube
3 x 0.25 mm
2
Oil pressure switch
D160.x6.1 3 and 4
X-ray tube
Iontomat cable
SDM connection
D100.X28
I.I. bracket D100.X1
Tab. 9
K32
K31
Fig. 9
Fig. 10
F
H
L