Connections with POLYDOROS LX 30
3 - 17
Siemens AG
RXL2-120.812.01
Page 17 of 20
LITHOSTAR Multiline
Medical Solutions
Rev. 02
10.04
CS PS 24
System
Connecting the options
Pilot lights and radiation protection indicators
At the customer's request or when country-specific regulations require (e.g. DHHS), the
following pilot lights and displays can be connected in the POLYDOROS LX to D160.X6
(
/Fig. 14):
•
Door contact for blocking the exposure release. If a door contact is required, it must be
connected in series with the oil pressure switch.
(Have the door contact activated e.g. by the examination room door.)
•
External radiation indicator lamp
•
Remove the corresponding jumper and connect the function according to wiring diagram
X2169-11. If necessary, pull the left part of the D160.X6 plug out from the side.
Generator - Periphery
Type of Cable
Function
Generator
ok
Power line cable
3/N/Gnd
System
power line connection
M16.
F1, F2, F3, N,
Main distributor
L1, L2, L3, N,
2 x 0.75 mm
2
Radiation protection
9)
M16.D160.X6
Door contact
2 x 0.75 mm
2
Radiation buzzer
Radiation display
9)
M16.D160.X6
Radiation indicator
lamps
Tab. 10
9)
Option
Fig. 14
J
K
K
K