LITHOSTAR Multiline
RXL2-120.812.01
Page 12 of 14
Siemens AG
System
Rev. 02
10.04
CS PS 24
Medical Solutions
2 - 12
System Installation
Installing the TV camera on the 23" I.I.
2
•
Remove the housing (1/Fig. 18) from the TV camera.
•
Remove the cover of the I.I. optics (black lid).
•
Remove the set screws on the longer sides of the TV camera.
•
Place the TV camera on the I.I. and mount it with the two set screws (
/Fig. 18).
•
Plug in the two connection cables on the TV camera and secure them (Fig. 18).
•
Clamp the shielding of the camera cable D115.X2 under the clamp.
•
Reinstall the TV camera housing (1/Fig. 18) but do not secure it.
•
Connect the protective conductor to the white cover (2/Fig. 18).
•
Place the white cover (2/Fig. 18) over the TV camera and secure it together with the
housing.
•
Use the white plastic plugs to cover the holes.
Installing the TV camera on the 33" I.I.
2
•
The camera has been pre-installed at the factory.
Checking the temperature of the image intensifier
2
•
Check the temperature indicators of the image intensifier according to the installation
instructions regarding temperature checks with SIRECON units and OPTILIX tubes in
RXD0-000.038.01... The results of the evaluation are recorded in the image quality
protocol (IQC) during start-up.
1
2
Fig. 18