LITHOSTAR Multiline
RXL2-120.812.01
Page 2 of 20
Siemens AG
System
Rev. 02
10.04
CS PS 24
Medical Solutions
3 - 2
Connections with POLYDOROS LX 30
Fig. 2
Connections and cable locations in the A12
electronics cabinet
3
(green cable harnesses)
•
Fasten the 3 cable harnesses (green) with the cable clip on the rail that
is installed at the bottom of the electronics cabinet.
•
Secure the braided shield of the zipper hoses to the locations provided.
•
Route the cables in the A12 electronics cabinet as shown in the adjacent
Fig. 2 and connect them according to table 1 and 3.
The numbers or letters (e.g.
) are located in the cable lists on the
following pages.
•
Prior to routing the system cables, remove the EMC shielding plate at
N11 D11.
right
- cable harness,green, from control console
centre - cable harness,green, from unit (K3, K4....)
left
- cable harness, green, from unit (K8, K9....)