By knowing the time between the first and second reflections, along with knowing the
upper layer dielectric constant, the thickness of the upper layer can be determined. In
order to properly process the reflected signals, the SITRANS LG200 is specified for those
applications where the thickness of the upper layer is greater than 2 inches. The
maximum upper layer is limited to the length of the Model 7ML1301-6 probe, which is
available in lengths from 60 to 610 cm (24 to 240").
Emulsion Layers
As emulsion (rag) layers can decrease the strength of the reflected signal, the SITRANS
LG200 should only be used in applications that have clean, distinct layers. Contact the
factory for application assistance.
Maximum Process Temperature Chart
Note:
The charts below are for guidance only: