Page 8
SITRANS LG200 – INSTRUCTION MANUAL
7ML19985KA02
Specificat
ions
• LCD temperature range
-20 to +70 °C (-5 to +160 °F)
• humidity
0 to 99%, non-condensing
Measurement accuracy
• linearity
1
<12.7 mm (0.5")
• measured error
Upper layer: ± 25.4 mm (1")
Interface layer: ± 25.4 mm (1") (clean distinct interface
required)
• resolution
±2.5 mm (0.1")
• repeatability
<12.7 mm (0.
5
")
• hysteresis
<12.7 mm (0.5")
Update time
< 1 second
Influence of amb. temperature +0.02% of probe length/°C
Upper layer dielectric
1.4 to 5.0
Interface layer dielectric
>15
Electromagnetic compatibility
2
Meets CE requirements (EN 61000-6-2/2001,
61000-6-4/2001)
Input
• measured variable
level, determined by time-of-flight of guided radar pulse
from transmitter to product surface and back
• zero and span
15 cm to 22.86 m (6" to 75 ft)
• user interface
3 button menu-driven data entry and system security
• digital communication
HART Version 5.x compatible
Output
• mA
Analog: 4 to 20 mA with HART digital signal
• range
Analog: 3.8 to 20.5 mA usable
• resolution
Analog: 0.01 mA
Digital: 2.5 mm (0.1")
• start-up current
4 mA
1.
Top 610 mm (24") of Model 7ML1302-1 probe: 30 mm (1.2"). Specification for top 1.22 m (48") will be
application dependent.
2.
Single and Twin Rod probe versions should only be used in metallic vessels or stilling wells for
compliance with CE regulations (where applicable)
Содержание 7ML1301-1
Страница 1: ...Instruction Manual February 2009 LG200 sitrans ...
Страница 10: ...Page 6 SITRANS LG200 INSTRUCTION MANUAL 7ML19985KA02 SITRANS LG200 Notes ...
Страница 59: ...7ML19985KA02 SITRANS LG200 INSTRUCTION MANUAL Page 55 SIMATIC PDM ...
Страница 60: ...Page 56 SITRANS LG200 INSTRUCTION MANUAL 7ML19985KA02 SIMATIC PDM Notes ...
Страница 96: ...Page 92 SITRANS LG200 INSTRUCTION MANUAL 7ML19985KA02 Hazardous Installation ...