Firmware Version
Product Identification firmware version for SITRANS LG200
Display Factory Parameters (Disp Fact)
Select Yes to display factory parameter menus. After five minutes of keypad inactivity,
reverts to No. For authorized personnel only.
Diagnostic and Factory Parameters (Factory Authorized
Personnel Only)
History
(factory authorized personnel only)
View history of exceptions and their duration since the most recent power on.
Run Time
(factory authorized personnel only)
View the value of the elapsed time counter since power on or a History reset.
History Reset
(factory authorized personnel only)
Clear exception history and time counter by pressing enter (similar to Safety Zone Alarm
Reset); Superuser password required.
Fiducial Ticks (Fid Ticks)
(factory authorized personnel only)
Time of flight in counts from reference point to measured point. This value can be
converted from equivalent time to real distance via calculations.
Fiducial Spread (Fid Sprd)
(factory authorized personnel only)
Variability measure of Fiducial Ticks.
Fiducial Type (Fid Type)
(factory authorized personnel only)
Default: negative. Select from positive or negative (only available on selected probe
models); Superuser password required.
Fiducial Gain (Fid Gain)
(factory authorized personnel only)
Default: 90. Amount of gain applied to fiducial signal; Superuser password required.
Window (Window)
(factory authorized personnel only)
Default: 204. Used to adjust Fiducial Ticks. Typical values are between 200 and 212 for
integral units and between 150 and 165 for remote units; factory password required.
Conversion Factor (Conv Fct)
(factory authorized personnel only)
Default: 985. Calibration parameter. This value equals (mm/tick)*2048. The nominal value
is 963. This is the slope of the straight line measurement equation; factory password
required.
Values (View
only)
Product identification and firmware version displayed
Values
DispFact: Yes or No
Содержание 7ML1301-1
Страница 1: ...Instruction Manual February 2009 LG200 sitrans ...
Страница 10: ...Page 6 SITRANS LG200 INSTRUCTION MANUAL 7ML19985KA02 SITRANS LG200 Notes ...
Страница 59: ...7ML19985KA02 SITRANS LG200 INSTRUCTION MANUAL Page 55 SIMATIC PDM ...
Страница 60: ...Page 56 SITRANS LG200 INSTRUCTION MANUAL 7ML19985KA02 SIMATIC PDM Notes ...
Страница 96: ...Page 92 SITRANS LG200 INSTRUCTION MANUAL 7ML19985KA02 Hazardous Installation ...