Therefore avoid:
● Air in pipe.
● Installation at the highest point in pipe system.
● Installation in vertical pipes with free outlet.
Partially filled pipes
For partially filled pipes or pipes with downwards flow and free outlet, sensor must be mounted
in a U-tube.
4.1.2
Orienting the sensor
MAG 8000 CT with OIML R49:2013 and MID approval DN 50 to DN 600 can only be installed
horizontally.