Output B as call-up output

Call up output

Call up reset

Call up status

On - Off

When output B is configured as "call-up", the output is activated by an alarm condition and 

remains on until it is reset via meter display key or communication interface.
A new alarm will not activate a "call-up" function if the "call-up" function is still active from a 

previous alarm.

Note
Like alarm output, call-up output inverts to a pulse output providing a call-up if power 

disappears or cable connection is interrupted.

MAG 8000
When output B is configured as an "call-up" output, it will follow the internal alarms that were 

previously chosen in the Alarm Configuration List.

Note
Call-up output is inverted to a pulse output providing an alarm if power disappears or cable 

connection is interrupted.

Regional factory default settings

DN
mm

size
(inch)

Pulse width
ms

Amount per 

pulse
m

(Europe)

Amount per 

pulse
Gallons (USA)

Amount per 

pulse
ML (Australia)

25, 40, 50

(1", 1½", 2")

50

0.01

1

0.001

65, 80, 100, 125, 150 

(2½", 3", 4", 5", 6")

50

0.1

10

0.001

200, 250, 300, 350,
400, 450, 500

(8", 10", 12", 14",
16", 18", 20")

50

1

100

0.01

600, 700, 800, 900,
1000, 1050, 1100, 1200

(24", 28", 30", 32", 36",
40", 42", 44", 48")

50

10

100

0.01

Technical data
10.6 Output characteristics

SITRANS FM MAG 8000

100

Operating Instructions, 03/2020, A5E03071515-AH

Содержание 7ME6810

Страница 1: ...tion 1 Safety notes 2 Description 3 Installing Mounting 4 Connecting 5 Commissioning 6 Operating 7 Service and maintenance 8 Diagnostics and Troubleshooting 9 Technical data 10 Product documentation and support A Spare parts Accessories B Flow Tool C Qualification certificate D Unit conversion table E Parameter lists F Sizing sensor G Features H ...

Страница 2: ...el are those who based on their training and experience are capable of identifying risks and avoiding potential hazards when working with these products systems Proper use of Siemens products Note the following WARNING Siemens products may only be used for the applications described in the catalog and in the relevant technical documentation If products and components from other manufacturers are u...

Страница 3: ...allation in hazardous area 12 3 Description 13 3 1 System components 13 3 2 Operating principle 13 3 3 Design 13 3 4 Benefits 15 4 Installing Mounting 17 4 1 Sensor installation 18 4 1 1 Locating the sensor 18 4 1 2 Orienting the sensor 20 4 1 3 Mounting the sensor 22 4 2 Potential equalization 25 4 3 Grounding 26 4 4 Cathodic protected pipes 28 4 5 Potting and direct burial 28 4 6 Transmitter ins...

Страница 4: ...nit selection 56 6 5 Output configuration 56 6 6 Data protection 58 7 Operating 59 7 1 Operation via key and display 59 7 2 Display symbols 60 7 3 Default display information and accessible display menus 61 7 4 Operator menu 64 7 5 Internal data handling 70 7 6 Battery powered operation 71 8 Service and maintenance 77 8 1 Maintenance 77 8 2 MAG 8000 service guidelines 77 8 3 Replacing transmitter ...

Страница 5: ...8000 7ME6810 and MAG 8000 CT 7ME6820 108 10 12 Dimension drawings 109 A Product documentation and support 115 A 1 Product documentation 115 A 2 Technical support 116 B Spare parts Accessories 117 B 1 Ordering of spare parts 117 C Flow Tool 119 C 1 Flow Tool 119 C 2 Initial commissioning via Flow Tool 119 C 3 Configuring the device 120 C 4 Setting the basic parameters 122 C 5 Unit selection 126 C 6...

Страница 6: ...ion table 139 F Parameter lists 141 F 1 1 99 141 F 2 100 199 142 F 3 200 299 143 F 4 300 399 146 F 5 400 499 149 F 6 500 599 150 F 7 600 799 152 F 8 800 899 156 G Sizing sensor 161 G 1 Sizing table DN 25 to 1200 1 to 48 161 H Features 163 Index 169 Table of contents SITRANS FM MAG 8000 6 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 7: ... engineers 1 2 Items supplied SITRANS F M MAG 8000 Standard MAG 8000 CT Siemens Process Instrumentation docu mentation disk containing certificates and manuals Note Scope of delivery may vary depending on version and add ons Make sure the scope of delivery and the information on the nameplate correspond to your order and the delivery note Inspection 1 Check for mechanical damage due to possible im...

Страница 8: ...ered items for visible damages 2 Report any claims for damages immediately to the shipping company 3 Retain damaged parts for clarification 4 Check the scope of delivery by comparing your order to the shipping documents for correctness and completeness WARNING Using a damaged or incomplete device Risk of explosion in hazardous areas Do not use damaged or incomplete devices 1 5 Security information...

Страница 9: ...ent parts should be returned in their original packaging If the original packaging is no longer available ensure that all shipments are properly packaged to provide sufficient protection during transport Siemens cannot assume liability for any costs associated with transportation damages NOTICE Insufficient protection during storage The packaging only provides limited protection against moisture a...

Страница 10: ...Introduction 1 7 Notes on warranty SITRANS FM MAG 8000 10 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 11: ...Electrical Code NEC NFPA 70 USA Canadian Electrical Code CEC Canada Further provisions for hazardous area applications are for example IEC 60079 14 international EN 60079 14 EU 2 3 Conformity with European directives The CE marking on the device symbolizes the conformity with the following European directives Electromagnetic compatibili ty EMC 2014 30 EU Directive of the European Parliament and of...

Страница 12: ... hazardous substances in electri cal and electronic equipment The applicable directives can be found in the EU declaration of conformity of the specific device 2 4 Lithium batteries Lithium batteries are primary power sources with high energy content designed to provide the highest possible degree of safety WARNING Potential hazard Lithium batteries may present a potential hazard if they are abuse...

Страница 13: ...G 8000 is a microprocessor based water meter with graphical display and key for optimum customer operation and information on site The transmitter drives the magnetic field in the sensor evaluates the flow signal from the sensor and calculates the volume passing through Its intelligent functionality information and diagnostics ensure optimum meter performance and information to optimize water supp...

Страница 14: ...ndard compact Remote Figure 3 3 MAG 8000 Standard remote Custody transfer Compact and remote versions Figure 3 4 MAG 8000 CT compact version Description 3 3 Design SITRANS FM MAG 8000 14 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 15: ...tivity class Conformity with country specific directives Figure 3 5 MAG 8000 CT nameplate example 3 4 Benefits Simple placement of the meter bury the meter underground or in a floating chamber The IP 68 NEMA 6P design is unaffected by meter position or in line piping stresses and there is no requirement for filters Minimal pressure loss an unrestricted flow tube ensures minimal pressure loss even ...

Страница 16: ...Description 3 4 Benefits SITRANS FM MAG 8000 16 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 17: ... not exceeded General information This chapter describes how to install the water meter in the compact version as well as in the remote version 87L130 10 Figure 4 1 Compact installation Figure 4 2 Remote installation The installation consists of two steps 1 Sensor installation 2 Transmitter installation remote version only SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 17 ...

Страница 18: ...ction as indicated on label If process flow direction is opposite of flow direction indicated on sensor label forward flow rates can be restored via parameter 327 if factor is adjusted to 1 MAG 8000 CT is approved for bidirectional measurement ORZ GLUHFWLRQ Inlet and outlet condition To achieve most accurate flow measurement it is essential to have certain straight inlet and outlet pipe lengths as...

Страница 19: ...pe downstream from the sensor For optimal measurement performance it is recommended to follow the installation guidelines for the MAG 8000 standard water meter MAG 8000 recommended MAG 8000 CT DN 50 to DN 300 MAG 8000 CT DN 350 to DN 600 Minimum straight inlet pipe length 5x Di 0x Di 3x Di Minimum straight outlet pipe length 3x Di 0x Di 3x Di Sensor must be completely full of liquid Installing Mou...

Страница 20: ...rtially filled pipes or pipes with downwards flow and free outlet sensor must be mounted in a U tube 4 1 2 Orienting the sensor MAG 8000 CT with OIML R49 2013 and MID approval DN 50 to DN 600 can only be installed horizontally Horizontal pipes Installing Mounting 4 1 Sensor installation SITRANS FM MAG 8000 20 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 21: ...e Physical installation of battery pack may influence battery capacity Optimal battery capacity is achieved with battery pack in an upright position Installation examples marked with dotted cross will affect battery capacity Vertical pipes MAG 8000 Recommended installation is in a vertical inclined pipe to minimize wear and deposits in sensor Installation in large pipes MAG 8000 The water meter ca...

Страница 22: ...pressure drop of 2 9 mbar 0 04 psi 4 1 3 Mounting the sensor 1 Install gaskets 2 Ensure connection flange has a smooth surface and is in line with sensor Gaskets are recommended but not included in water meter delivery Installing Mounting 4 1 Sensor installation SITRANS FM MAG 8000 22 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 23: ...luid Hardness should be maximum 75 Shore A Vibrations In installations with strong vibrations remote installation of the transmitter is recommended Alternatively mitigate vibrations by installing pipe support in close proximity to the flowmeter Figure 4 3 Avoid vibrations Figure 4 4 Pipe support variant 1 Installing Mounting 4 1 Sensor installation SITRANS FM MAG 8000 Operating Instructions 03 202...

Страница 24: ...ter and mating flanges All bolts are galvanized and adequately lubricated Flanges are made of carbon steel Water meter and mating flanges are correctly aligned Torques for 7ME6810 and 7ME6820 Nominal size PN 10 PN 16 PN 40 Class 150 AWWA Mm inch Nm f lbs Nm f lbs Nm f lbs Nm f lbs Nm f lbs 25 1 N A N A N A N A 10 7 7 5 N A N A 40 1 N A N A N A N A 16 12 9 7 N A N A 50 2 N A N A 25 18 N A N A 25 18...

Страница 25: ... A N A N A N A N A N A N A 240 177 800 32 274 202 386 285 N A N A N A N A 260 192 900 36 288 213 408 301 N A N A N A N A 240 177 1000 40 382 282 546 403 N A N A N A N A 280 207 1050 42 N A N A N A N A N A N A N A N A 280 207 1100 44 N A N A N A N A N A N A N A N A 290 214 1200 48 395 292 731 539 N A N A N A N A 310 229 4 2 Potential equalization Liquid potential equalization or grounding is accomp...

Страница 26: ...ensor body is noise free Note Grounding connector rotated by 90 For sensor sizes between DN 25 1 and DN 300 12 the grounding connector on the sensor is rotated by 90 Metal pipes Connect straps to both flanges with 6 mm 1 4 screws Metal pipes Grounding straps rotated by 90 on DN 350 to DN 1200 Grounding straps are part of delivery and pre mounted on water meter Installing Mounting 4 3 Grounding SIT...

Страница 27: ...astic pipes Use straps for metal pipe and grounding rings for plastic pipe Metal pipe Plastic pipe Bonding grounding straps grounding rings and straps are not included in delivery Note All straps or grounding wires must be 12 AWG or heavier copper wire and connected with 6 mm screws Installing Mounting 4 3 Grounding SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 27 ...

Страница 28: ...ons Do not pot meter before electrical connections have been made Meter is rated IP68 NEMA 6P from the factory as standard If cable glands are used IP68 NEMA 6P enclosure rating is obtained by potting transmitter bottom with Sylgard potting kit Otherwise only an IP67 NEMA 4 rating is obtained Sealing of transmitter 1 Select the proper gland size to fit installed cable size 2 Mount O ring properly ...

Страница 29: ...or is protected to IP68 NEMA 6P and can be buried The use of pea gravel at least 300 mm 12 inches all around sensor is mandatory to provide some drainage and to prevent dirt from solidifying on sensor It also helps to locate the sensor should excavation be necessary Before covering pea gravel with earth use electrical cable identification tape above gravel Run remote sensor cable through a plastic...

Страница 30: ...ing Figure 4 7 Pipe mounting vertical Figure 4 8 Pipe mounting horizontal 4 7 MAG 8000 CT Calibration sealing has been carried out at calibration Installing Mounting 4 7 MAG 8000 CT SITRANS FM MAG 8000 30 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 31: ...y be broken by an authorized person with the acceptance and under direction of the local authorities Figure 4 9 MAG 8000 CT verification sealing A 4 7 1 2 User sealing Note User sealing has to be done after commissioning by an authorized person 1 Lead one or two wires through the sealing bolts 2 Seal the ends of each wire Installing Mounting 4 7 MAG 8000 CT SITRANS FM MAG 8000 Operating Instructio...

Страница 32: ...00 2 to 24 only in horizontal installation Note Straight pipe upstream and downstream DN 50 to DN 300 requires 0x D of straight pipe upstream and downstream from the sensor DN 350 to DN 600 requires 3x D of straight pipe upstream and 3x D downstream from the sensor Other restrictions may apply see certificate Installing Mounting 4 7 MAG 8000 CT SITRANS FM MAG 8000 32 Operating Instructions 03 2020...

Страница 33: ...ion Output A Output B The 3 6 V DC battery connector male and pulse connection terminals are placed in the right side of the PCB board see figure Connection for add on interface modules is placed on the left side HL Hardware lock key connection V Verification button enables verification mode for 4 hours To configure the outputs via PDM see Output configuration Page 56 To configure the outputs via ...

Страница 34: ...ected to the power supply If the housing is under voltage power supply the cover may be unscrewed by qualified personnel only WARNING Mains supply from building installation Class II A switch or circuit breaker max 15 A must be installed in close proximity to the equipment and within easy reach of the operator It must be marked as the disconnecting device for the equipment Connecting 5 1 General s...

Страница 35: ...RADIUS of 45 mm 1 8 inches otherwise there s a risk of damaging the cable Save dust covers for future use and protection 3 Ensure connectors are clean 4 Ensure connectors are fastened securely to achieve a good connection and watertight seal Note If dirt enters connector ends use plain water for cleaning Ensure connectors are completely dry before making connections For MAG 8000 CT user sealing se...

Страница 36: ...be connected to male connector 3 6 V DC on PCB board Battery backup input Male battery connector with black and red wires black wire is ground Male battery connector has to be connected to female connector on backup battery Functional ground Black wire with terminal must be connected to the device encapsulation with a screw Mains power supply has to be connected to a switch near water meter accord...

Страница 37: ...apsulation with a screw Cable installation Choose the correct glands for the selected cable type see Accessories for glands selection Ensure shield is mounted under cable clamps do not make pig tails Note Mains or line powered PUR cable no shield has to be mounted under cable clamps All cable glands have to be sufficiently tightened to ensure IP rating Connecting 5 3 Power supply SITRANS FM MAG 80...

Страница 38: ...up see Commissioning Page 56 Pulse output is not polarized and can be connected for positive or negative logic Pull up down resistor R is selected in relation to power supply voltage V and with a max current I of 50 mA Note Pulse output isolation The MAG 8000 pulse output has a functional isolation only and must therefore be connected to equipment complying with the Low Voltage Directive in order ...

Страница 39: ...lse count of an external measured value display is certified by the national metrological authority Pulse B function is only allowed to be set to Alarm or call up in MAG 8000 CT with MID 001 approval 5 5 Communication modules RS 232 connection diagram Common Connect shield to encapsulation Shield must be connected to protective ground Common A Modbus over serial line cable must be shielded At one ...

Страница 40: ...d must always be connected to MAG 8000 encapsulation using cable clamp as shown in figure in section cable installation see Power supply Page 36 Bus termination All RS 485 based networks must be terminated correctly to function properly A termination must be placed at each end of segment The Modbus RTU module can add a 120 ohm termination if a jumper is placed beside terminals in position ON Termi...

Страница 41: ... terminal 93 WARNING Unshielded wire It is important that unshielded wire does not touch any metal parts of MAG 8000 housing Other radio interface cable has to be a 3 wire cable with a shield connected to MAG 8000 housing mounting cable shield is shown to the right GND Data 129 10 10 Power clock Connect shield to encapsulation Radio interface 128 10 10 Connecting 5 5 Communication modules SITRANS ...

Страница 42: ...iemens com flowdocumentation www siemens com flowdocumentation Information on the operation of communication modules For information on how to operate the various communication modules see 3G UMTS add on module Operating Instructions https support industry siemens com cs ww en view 109751612 Modbus RTU Operating Instructions http support automation siemens com WW view en 54091509 Connecting 5 6 Co...

Страница 43: ...ic pdm SIMATIC PDM monitors the process values alarms and status signals of the device It allows you to display compare adjust verify and simulate process device data also to set schedules for calibration and maintenance For information on for example how to install and integrate devices commission the software see Operating Manual Help for SIMATIC PDM The manual is delivered with SIMATIC PDM soft...

Страница 44: ...ce 5 Save files to your computer in an easily accessed location 6 Launch SIMATIC PDM Device Integration Manager From the File menu click Read device descriptions from compressed source 7 Browse to the zipped EDD file select and open it 8 Use the Integration function to integrate the EDD into the device catalog The EDD is now accessible via SIMATIC Manager This chapter describes how to install the ...

Страница 45: ...where the file was downloaded and saved The PDM will show a tree with the Software options Click on the Siemens checkbox to select all options Click OK to install the new EDD files Commissioning 6 2 Initial commissioning via SIMATIC PDM SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 45 ...

Страница 46: ...e the same only the views on the PC screen differ Setting up meter 1 Click on File 2 Select New 3 Assign a new project name in the Name box e g MAG8000 4 In Process Device Network View right click on Net select Insert New Object MODBUS net Commissioning 6 2 Initial commissioning via SIMATIC PDM SITRANS FM MAG 8000 46 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 47: ...5 Right click on MODBUS net select Object Properties Commissioning 6 2 Initial commissioning via SIMATIC PDM SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 47 ...

Страница 48: ...8000 is 19 200 baud If the IrDA adapter is connected properly a small IrDA icon will appear in the Windows notification area When placing mouse at this icon the device information will be shown e g MAG8000 SN1234 is in range Commissioning 6 2 Initial commissioning via SIMATIC PDM SITRANS FM MAG 8000 48 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 49: ...ow click on Flow Electromagnetic Siemens AG MAG8000 9 Select the proper device MAG 8000 Advanced MAG 8000 Basic MAG 8000 CT Advanced or MAG 8000 CT Basic depending on transmitter version to be controlled Commissioning 6 2 Initial commissioning via SIMATIC PDM SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 49 ...

Страница 50: ...click created device e g MAG8000 Advanced A new window will pop up for device parameter configuration Commissioning 6 2 Initial commissioning via SIMATIC PDM SITRANS FM MAG 8000 50 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 51: ... Settings to specify whether you are permanently registered as a Maintenance engineer or Specialist and or whether to disable the display of the User dialog box Refer to SIMATIC PDM help manual for more details of Maintenance engineer and Specialist Proceed as follows Click the appropriate button Maintenance engineer or Specialist If the option Specialist is selected enter the password in the Pass...

Страница 52: ...off line data which are not stored in the MAG 8000 yet whereas black texts show the actual meter data Each adjustable parameter has a Default Value a Minimum Value and a Maximum Value which are prompted by right clicking on the parameter More information will be available when the Help button is clicked Commissioning 6 3 Setting the basic parameters SITRANS FM MAG 8000 52 Operating Instructions 03...

Страница 53: ...marked alarms enabled Click on Transfer to instantly synchronize alarm configurations between the PC and the device Figure 6 1 Alarm status marked alarms are enabled Select Device Download to Device to download all the changed parameters to the device Commissioning 6 3 Setting the basic parameters SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 53 ...

Страница 54: ...right click on the PDM object to be exported select Export In the subsequent dialog box specify the type of export as well as the name and location where the PDM file containing the exported data is to be saved Confirm the entries by clicking OK Commissioning 6 3 Setting the basic parameters SITRANS FM MAG 8000 54 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 55: ... showing export progress After the export is completed the status Result is displayed See also www siemens com flow www siemens com flow Commissioning 6 3 Setting the basic parameters SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 55 ...

Страница 56: ... be stored into the transmitter via Download to Device 6 5 Output configuration Pulse output can be configured as volume pulse alarm or call up Default factory setting is with output A enabled for forward flow and output B for alarm output Output configuration in SIMATIC PDM Select Output The software window shows the parameters for output configuration Each parameter has its own guideline for sel...

Страница 57: ...See Output characteristics Page 98 for more information on how the outputs work Commissioning 6 5 Output configuration SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 57 ...

Страница 58: ...bration date 323 Calibration factor 325 Sensor offset 332 Max sensor excitation frequency Additional protected parameters MAG 8000 CT Parameter no Parameter name 101 Totalizer 1 102 Totalizer 2 103 Net totalizer since FW 3 11 303 Operation excitation frequency 305 Decimal point place 310 Flow direction totalizer 1 312 Flow direction totalizer 2 327 Adjustment factor 328 Low flow cut off 332 Empty ...

Страница 59: ...mation see menu overview Page 64 If key is not pressed for 10 minutes display will time out and return to default configured operator menu Key There are three different ways the interface key will respond to being pressed 1 A brief press less than 2 seconds will advance screen to next index or menu 2 A short press 2 to 5 seconds will enter a menu or escape menu selection 3 A long press more than 5...

Страница 60: ...ch key s self calibration once the MAG 8000 transmitter enclosure is removed and mounted again There will be no response from the touch key during self calibration 7 2 Display symbols Top area of display shows status bar 6WDWXV EDU FWXDO WDULII DFFRXQW DQG UHVHW LQGLFDWLRQ XUUHQW DFWLYH PSW SLSH DFWLYH 0DLQV SRZHU VXSSO DWWHU SRZHU VXSSO DQG HQHUJ VWDWXV Figure 7 3 Status bar Status information sy...

Страница 61: ...H LQGLFDWLRQ HIDXOW RSHUDWRU PHQX 0HWHU LQIRUPDWLRQ PHQX 6HUYLFH PHQX DWD ORJJHU PHQX 6WDWLVWLFV PHQX DGYDQFHG YHUVLRQ RQO 7DULII PHQX DGYDQFHG YHUVLRQ RQO 0HQX EDU Figure 7 4 Menu bar The menu bar icons indicate actual selected menu and related index for selected information Display overview shows relation between menu index and information Only operator menu has information and functions that ca...

Страница 62: ... accessible display menus with selection of one or more of the following menus Operator menu Meter info menu Service menu Data logger menu Statistic menu advanced version only Revenue menu advanced version only Operating 7 3 Default display information and accessible display menus SITRANS FM MAG 8000 62 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 63: ...Operating 7 3 Default display information and accessible display menus SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 63 ...

Страница 64: ...7 4 Operator menu The operator menu consists of several indexes described in the following Operating 7 4 Operator menu SITRANS FM MAG 8000 64 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 65: ... 3DULW 6WRS ELW Figure 7 5 Menu overview Operating 7 4 Operator menu SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 65 ...

Страница 66: ...menu Totalizer 2 Flow volume totalizer 2 factory configured for reverse flow A negative value indicates reverse flow calculation For MAG 8000 the value of totalizer 2 can be reset to zero or set to any value desired via PDM or Flow Tool example replacing an existing old meter Index 3 Flow rate Figure 7 8 Operator menu Flow rate Index 3 shows actual flow rate If a negative value is indicated flow i...

Страница 67: ...warning 5 leakage warning L and empty pipe warning E Faults 1 to 4 affect meter performance and remain active until alarm condition disappears Faults 5 to d are warnings that will disappear when alarm condition has been corrected and they are reset via communication interface Fault evaluation and service guidelines are made in service section After all faults have disappeared display shows total h...

Страница 68: ...ts are configurable A Reverse high flow rate warning alarm limits are configurable Meter disables measurement to reduce power consumption during fatal faults Index 5 Customer totalizer Figure 7 10 Operator menu Totalizer For the CT version the totalizer is neither configurable nor resttable Display test Figure 7 11 Operator menu Display test All segments of display are alternately flashed on and o...

Страница 69: ...ed in parameter 130 in advance Index 0 when active Call up reset RQJ SUHVV IRU DFFHSW UHVHW LQIRUPDWLRQ IXQFWLRQ DLW IRU V PERO WR IODVK VHF 5HVHW DFFHSWHG 5HVHW SRVVLEOH Figure 7 13 Operator menu Call up reset Call up reset window index 0 is only shown when call up function is activated r indicates that it can be reset by a long press on key When releasing key while r is flashing call up function...

Страница 70: ...data handling Meter status Meter status parameter 120 gives a fast indication of reliability of revenue data Operating 7 5 Internal data handling SITRANS FM MAG 8000 70 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 71: ... the next logging is made and so on The consumption alarm indicates that the actual consumption on totalizer 1 is above or below the consumption limits Note Invalid log date If the log date 00 00 2000 is shown in the device display the log period was never set 7 6 Battery powered operation The MAG 8000 Standard 7ME6810 and MAG 8000 CT 7ME6820 are factory configured for 7 years of typical operation...

Страница 72: ...ng elements and the influence of temperature changes Consumption and operation time calculation Battery operation time depends on connected battery pack as well as operation conditions of meter Every 4 hours the advanced power management system calculates the real power consumption and remaining operation capacity Power consumption calculation includes flow measurement meter dialog communication a...

Страница 73: ...00 to 1200 28 to 48 7 years 4 years 2 years 1 year 1 month N A N A Four D Cell battery 66 Ah External bat tery pack DN 25 to 150 1 to 6 15 years 15 years 14 years 86 months 16 months 7 months 4 months DN 200 to 600 8 to 24 15 years 13 years 8 years 44 months 7 months 3 months N A DN 700 to 1200 28 to 48 14 years 9 years 5 years 2 years 3 months N A N A Typical operation time of 7 years is based on...

Страница 74: ...y of one D cell battery in relation to discharge current and temperature Note Installation orientation of battery pack may influence battery capacity Optimal battery capacity is achieved with battery pack in an upright position Operating 7 6 Battery powered operation SITRANS FM MAG 8000 74 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 75: ...e located in Alarm group e g Battery limit parameter 206 which is the level at which low power alarm is activated generating an alarm or call up if configured as well as low power fault related parameters 230 to 234 When switching between battery power from the internal and the external battery packs Battery power parameter 507 must be adjusted to match the actual number of batteries connected Ope...

Страница 76: ...Operating 7 6 Battery powered operation SITRANS FM MAG 8000 76 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 77: ...ion proper installation and proper commissioning for the particular application This service guideline section indicates how to detect and solve the most common problems Meter and application problems are indicated by the alarm program via the main fault and warning symbol on the display and the comprehensive data logging and monitoring available via the communication interface Alarm monitoring in...

Страница 78: ...ation can be uploaded from the old meter or it can be read from the old meter product label 3 Order only a replacement PCB board The PCB board can only be ordered for basic or advanced versions with default settings When making the configuration on site the service mode must be selected in the Flow Tool SIMATIC PDM and the hardware key must be mounted on the PCB board to change important parameter...

Страница 79: ...lica gel bag 4 Replace O ring to ensure continued IP68 enclosure rating Check O ring for damage or deformity Smear O ring with acid free lubricating gel 5 Push locking tab and loosen strip 2OG 1HZ 1HZ 6 Remove battery pack with power still connected Service and maintenance 8 4 Battery replacement SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 79 ...

Страница 80: ...10 Press key within 6 seconds to reset battery i e operating time and remaining capacity when display shows 11 Mount top lid 12 Fasten screws to reassemble meter completely If necessary adjust time and date via software see Power up with battery reset date and time set up Note Re verification All sealed MAG 8000 CT meters must be re verified when sealings marked A have been broken The re verificat...

Страница 81: ...for 10 seconds Display will then show rESEt bAt indicating the option to reset internal battery power calculation To execute reset press key within 6 seconds If key is not pressed meter will proceed to set date set clock and finally normal operation mode If key is pressed within reset battery time display will indicate Accept to ensure that reset should take place Reset will take place only if key...

Страница 82: ...sed to 1 kHz and pulse width is set to 0 5 ms When verification mode is exited the previous pulse setting is restored Pulse widths other than 1 ms can be selected by storing new pulse values This setting remains when verification mode is exited Activation of verification mode Verification mode is enabled in one of the following ways Pressing verification button through hole in front screen Writing...

Страница 83: ...the device when transporting it could become loose as a result of vibrations and cause damage within the device Remove the batteries before returning the device Dispose of the batteries according to regulations or include them separately in the return consignment 8 8 Return procedure Enclose the bill of lading return document and decontamination certificate in a clear plastic pouch and attach it f...

Страница 84: ...d without a decontamination declaration will be cleaned at your expense before further processing Note Return of products with lithium batteries Lithium batteries are dangerous goods according to the Regulation of Dangerous Goods UN 3090 and UN 3091 Remove lithium batteries prior to shipment If the battery cannot be removed return the product according to the Regulation of Dangerous Goods with spe...

Страница 85: ...osal services Waste industrial batteries from our products are accepted back by Siemens and by the local Siemens representatives Please follow the return procedures Page 83 of Siemens or talk to your local Siemens partner Service and maintenance 8 9 Disposal SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 85 ...

Страница 86: ...Service and maintenance 8 9 Disposal SITRANS FM MAG 8000 86 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 87: ...nsulation test parameter 800 will reset alarm and start a new insu lation test During the 4 minute insulation test meas urement is disabled Alarm remains active until a new test period detects no failures 2 Coil current fault Error in the coil circuit Coil current has not set tled within specified pe riod A short circuit in the coil can NOT be detected Coils have been discon nected could be caused...

Страница 88: ...actual installation Alarm remains active until fault condition is resolved 7 PulseA over load Duty cycle of output A has exceeded maxi mum possible of 50 Wrong settings for out put A Change volume per pulse to a higher value see Technical data Page 98 for pulse selection Note Pulse A is limited to 100Hz The alarm remains active until output pulse rate drops below maximum pulse rate 9 Consumption I...

Страница 89: ...ontrol pin every 6 respectively 5 samples the sample frequency is 9600 Hz The impedance value is averaged over 100 measurements The electrode impedance A and B are measured in turns The unfiltered impedance value a fast warning indication is compared with a limit and the empty pipe warning is reported when it exceeds this limit but only if the detection is ON and if there is no overload failure Du...

Страница 90: ... overload this measurement is cancelled and a failure is reported During overload detection the coil current is held OFF and the flow value is forced zero Checking facilities The MAG 8000 and MAG 8000 CT are equipped with checking facilities of types P I and N in OIML R49 The automatic checks are performed without operator intervention Type P permanent checking facilities are automatic checks perf...

Страница 91: ... to any connected device or system WARNING Totalized values are changed during simulation and actual flow is NOT measured Simulation will stop automatically in 4 hours For MAG 8000 CT flow simulation is forbidden Diagnostics and Troubleshooting 9 3 Flow simulation SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 91 ...

Страница 92: ...Diagnostics and Troubleshooting 9 3 Flow simulation SITRANS FM MAG 8000 92 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 93: ...mbient 20 to 60 C 4 to 140 F 20 to 55 C 4 to 131 F Media 0 to 70 C 32 to 158 F 0 1 to 50 C 32 to 122 F Storage 40 to 70 C 22 to 158 F 40 to 70 C 22 to 158 F Enclosure IP68 NEMA 6P rating Cable glands mounted requires Sylgard potting kit to remain IP68 NEMA 6P otherwise IP67 NEMA 4 rating is obtained Factory mounted cable provides IP68 NEMA 6P rating Approvals Drinking water approvals NSF 61 cold w...

Страница 94: ...4 16 bar 232 psi AWWA C 207 28 to 48 PN 10 AS 4087 DN 50 to 1200 PN 14 DN 50 to 600 PN 16 AS 2091 1 Table D AS 2191 Table E Max excitation frequency Basic version Battery pow ered 1 15 Hz for sensor size DN 25 to 150 1 to 6 1 30 Hz for sensor size DN 200 to 600 8 to 24 1 60 Hz for sensor size DN 700 to 1200 28 to 48 1 15 Hz for sensor size DN 50 to 150 1 to 6 1 30 Hz for sensor size DN 200 to 600 ...

Страница 95: ... Transmitter Transmitter MAG 8000Standard 7ME6810 MAG 8000 CT 7ME6820 Installation Integral compact or remote with factory mounted cable in 5 10 20 or 30 m lengths with IP68 NEMA 6P connectors Connection is made at the transmitter bottom Material Top housing Stainless steel AISI 316 Wall mounting bracket Stainless steel AISI 304 Display and key Display 8 digits for main information Index menu and ...

Страница 96: ...e as Pulse volume Forward Reverse Forward net Reverse net Output B Programmable as Pulse volume Forward Reverse Forward net Reverse net Alarm Call up Pulse rate Pulse A Max 100 Hz Pulse B Max 50 Hz Pulse width 5 10 25 50 100 500 ms Communication IrDA Standard integrated infrared communication interface with MODBUS RTU pro tocol Add on modules RS 232 serial interface with MODBUS RTU Rx Tx GND point...

Страница 97: ...Length 3 m Resistance Sunlight and water Outer diameter 7 mm 0 28 Rated voltage 300 500 V AC Testing voltage 2000 V AC Temperature range Fixed laying Flexible application 40 to 90 C 40 to 194 F 30 to 80 C 22 to 176 F Bending radius Min 28 mm fixed installation Pulling force Max 200 N Output Female connector Backup battery Male connector 1 Lithium batteries are subject to special transportation reg...

Страница 98: ...rofile None Add on modules comply with MODBUS over serial line specification implementation guide v 1 0 modbus org 12 02 02 MODBUS application protocol specification v 1 1 modbus org 12 06 02 Isolation 500 V AC functional isolation of data signals and common Add on modules must be connected to equipment complying with Low Voltage Directive in order to be considered safe The isolation within the MA...

Страница 99: ... exceed the pulse rate of the output configuration table MAG 8000 CT The volume per pulse is defined in the MAG 8000 CT version to 0 001 or 0 01 or 0 1 or 1 m3 per pulse Output B as alarm output Alarm output Alarm status On Off No Error MAG 8000 When output B is configured as an alarm output it will follow the internal alarms that were previously chosen in the Alarm Configuration List Page 53 Note...

Страница 100: ...figured as an call up output it will follow the internal alarms that were previously chosen in the Alarm Configuration List Note Call up output is inverted to a pulse output providing an alarm if power disappears or cable connection is interrupted Regional factory default settings DN mm size inch Pulse width ms Amount per pulse m3 Europe Amount per pulse Gallons USA Amount per pulse ML Australia 2...

Страница 101: ...0035 0 018 4 623 0 000018 0 035 0 175 200 8 1000 0 0028 0 0056 0 028 7 338 0 000028 0 056 0 278 250 10 1600 0 0044 0 0089 0 044 11 741 0 000044 0 089 0 444 300 12 2500 0 0069 0 0139 0 069 18 345 0 000069 0 139 0 694 350 14 3463 0 0096 0 0192 0 096 25 412 0 000096 0 192 0 962 400 16 4523 0 0126 0 0251 0 126 33 190 0 000126 0 251 1 256 450 18 5725 0 0159 0 0318 0 159 42 010 0 000159 0 318 1 590 500 ...

Страница 102: ...over time the net pulse output indicates the bi directional totalizer as calculated internally The same principle applies for forward and reverse flow calculation By changing the status of the pulse output the internal pulse calculator will be reset Flow Net totalizer in meter display Bi directional Pulse output forward Uni directional mode Volume m3 Pulse output net forward Bi direction al mode V...

Страница 103: ...N150 6 m3 0 0 0 0 0 0 0 0 630 1 7ME 6820 4P DN200 8 m3 0 0 0 0 0 0 0 0 1000 1 7ME 6820 4V DN250 10 m3 0 0 0 0 0 0 0 0 1600 1 7ME 6820 5D DN300 12 m3 0 0 0 0 0 0 0 0 1600 1 Parameter 305 Decimal Point should be configured to 1 digit after decimal point in case Encoder interface module is selected in MAG 8000 or MAG 8000 CT otherwise proper reading from the device cannot be performed 10 7 Meter unce...

Страница 104: ... 300 12 results in a max 0 2 of rate 2 mm s uncertainty A calibration certificate is included with every sensor and calibration data is stored within the meter 1 2 ORZUDWH P V UURU s MAG 8000 Stand ard Standard calibra tion 0 4 of rate 2 mm s MAG 8000 Stand ard Extended calibra tion 0 2 of rate 2 mm s Calibration references conditions ISO 9104 and DIN EN 29104 Media temperature 20 C 5K 68 F 9 F Am...

Страница 105: ...l MAG 8000 CT program is type approved according to international water meter standard OIML R49 2013 The Custody Transfer program is approved as Class I and Class II for the sensor program from DN 50 to DN 600 at different Q3 and Q3 Q1 Q2 Q1 1 6 and follows standard OIML R49 specification O O 2 0 5 ODVV ODVV Table 10 1 OIML R49 2013 for Class 1 Size 50 2 350 14 400 16 450 18 500 20 600 24 R Q3 Q1 ...

Страница 106: ...ivalent to OIML R49 accuracy Class 2 according to Directive 2014 32 EU of the European Parliament and Council of February 26 2014 on measuring instruments MID Annex MI 001 in the sizes from DN 50 to DN 600 The MID certification is obtained as a module B and module D approvals according to the above mentioned directive Module B Type approval according to OIML R49 Module D Quality insurance approval...

Страница 107: ...8 8 1 25 2 3 125 5 7 875 12 5 20 20 31 25 50 50 Size 50 2 65 2 80 3 100 4 125 5 150 6 200 8 250 10 300 12 350 14 400 16 450 18 500 20 600 24 R Q3 Q1 160 Q4 m3 h 50 79 125 200 312 5 500 788 1250 2000 3125 5000 5000 7875 7875 Q3 m3 h 40 63 100 160 250 400 630 1000 1600 2500 4000 4000 6300 6300 Q2 m3 h 0 4 0 63 1 1 6 2 5 4 6 3 10 16 25 40 40 63 63 Q1 m3 h 0 25 0 394 0 62 5 1 1 563 2 5 3 938 6 25 10 1...

Страница 108: ... OIML R49 2013 Pressure PN 16 bar or PN 10 bar or PN 6 bar Power supply 3 6 V Lithium Battery 12 24 VAC VDC 115 230 VAC Compact Remote Max 30 m cable 10 11 The effect of temperature MAG 8000 7ME6810 and MAG 8000 CT 7ME6820 Metric Pressures in bar Flange spec Flange rating Temperature C 0 10 50 70 Sizes 25 mm 40 mm and 300 mm EN 1092 1 PN 10 10 0 10 0 9 7 9 4 PN 16 16 0 16 0 15 5 15 1 PN 40 40 0 40...

Страница 109: ... 580 561 547 ANSI 16 5 150 lb 286 286 280 261 Sizes 2 to12 EN 1092 1 PN 10 145 145 145 119 PN 16 145 232 232 191 ANSI 16 5 150 lb 145 286 286 235 10 12 Dimension drawings Meter dimensions Dimensions for MAG 8000 Standard 7ME6810 and MAG 8000 CT 7ME6820 Technical data 10 12 Dimension drawings SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 109 ...

Страница 110: ...450 17 7 450 N A N A 212 8 35 2 71 160 300 12 317 12 6 500 500 19 7 500 N A N A 265 10 43 2 88 198 350 14 369 14 6 550 550 21 7 550 N A N A 350 13 78 2 127 279 400 16 394 15 6 600 600 23 6 600 N A N A 400 15 75 2 145 318 450 18 425 16 8 600 600 23 6 600 N A N A 450 17 72 2 175 394 500 20 450 17 8 600 600 23 6 600 N A N A 500 19 68 2 225 494 600 24 501 19 8 600 600 23 6 600 N A N A 600 23 62 2 340 ...

Страница 111: ... 48 810 31 9 1200 1500 N A 1200 1200 47 2 1200 47 23 2 887 1996 1 For remote version the sensor weight is reduced with 2 kg 4 5 lb 2 See flange table Page 109 Remote version Figure 10 1 Dimensions in mm inch weight 3 5 kg 8 lbs Flange dimensions MAG 8000 7ME8610 and MAG 8000 CT 7ME6820 Dimensions mm Bolting Size D PCD T B Holes Bolts PN 10 200 340 295 24 22 8 M20 Technical data 10 12 Dimension dra...

Страница 112: ...5 185 145 20 18 8 M16 80 200 160 20 18 8 M16 100 220 180 22 18 8 M16 125 250 210 22 18 8 M16 150 285 240 24 22 8 M20 200 340 295 26 22 12 M20 250 405 355 29 26 12 M24 300 460 410 32 26 12 M24 350 520 470 35 26 16 M24 400 580 525 38 30 16 M27 450 640 585 42 30 20 M27 500 715 650 46 33 20 M30 600 840 770 52 36 20 M33 750 800 1015 950 33 24 M30 900 1125 1050 39 28 M36 1000 1255 1170 42 28 M39 1050 11...

Страница 113: ... 8 14 21 18 75 1 38 1 12 12 1 16 23 5 21 25 1 44 1 12 16 1 18 25 22 75 1 56 1 25 16 1 1 8 20 27 5 25 1 69 1 25 20 1 1 8 24 32 29 5 1 88 1 38 20 1 1 4 External battery pack and cable Note Physical orientation of battery pack may influence battery capacity Optimal battery capacity is achieved with battery pack in an upright position as shown The cable for the battery pack connection must be ordered ...

Страница 114: ...d 7ME6820 Figure 10 2 Flat ring Sizes DN 350 to 600 7ME6810 and 7ME6820 and DN 700 to 1200 7ME6810 W W Figure 10 3 Type C grounding ring Technical data 10 12 Dimension drawings SITRANS FM MAG 8000 114 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 115: ...dustry Online Support https support industry siemens com cs ww en sc 2067 Download the app to your mobile device and scan the device QR code Product documentation by serial number Using the PIA Life Cycle Portal you can access the serial number specific product information including technical specifications spare parts calibration data or factory certificates Entering a serial number 1 Open the PI...

Страница 116: ...ces at Service Support http www siemens com automation service support Contact If you have further questions about the device contact your local Siemens representative at Personal Contact http www automation siemens com partner To find the contact for your product go to all products and branches and select Products Services Industrial automation Process instrumentation Contact address for business...

Страница 117: ...t your ordering data is not outdated The latest ordering data is always available on the Internet SIOS catalog https support industry siemens com cs products dtp Catalog mfn ps pnid 17318 lc en US SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 117 ...

Страница 118: ...Spare parts Accessories B 1 Ordering of spare parts SITRANS FM MAG 8000 118 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 119: ...r The Device Drivers are related to meter version and is automatically selected in Automatic mode In Manual mode meter version is manually selected and version check is automatically made when data is uploaded or downloaded Data backup Meter information is stored in an internal data prom that secures data in case of power failure Information like totalizer 1 and 2 date and time and the statistic d...

Страница 120: ...Double click on Flow Tool icon on windows desktop to start program Note For on line communication make sure the IrDA symbol is visible in the toolbar status menu before starting Flow Tool C 3 Configuring the device This chapter describes how to set up meter for communication with PC The options in both Flow Tool and SIMATIC PDM software are the same only the views on the PC screen differ Flow Tool...

Страница 121: ...ght click on Project 2 Select New 3 Select Flow Meter 4 Name device 5 Select configuration type Automatic or Manual Flow Tool C 3 Configuring the device SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 121 ...

Страница 122: ...nloaded to the meter af terwards C 4 Setting the basic parameters Note No factory reset The device is delivered with factory settings that are not stored as default values Because no default values are present in the meter an automatic return to factory values is not possible Flow Tool C 4 Setting the basic parameters SITRANS FM MAG 8000 122 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 123: ...ritten printed or exported to a CSV file Only parameters data shown on white background can be changed Red texts are off line data which are not stored in the MAG 8000 yet whereas black texts show the actual meter data Each parameter has a description prompt advising how the parameter can be programmed and which settings are programmable The figure below shows alarm status with the marked alarms e...

Страница 124: ...o various functional groups with maximum 99 parameters included For a complete parameter overview see Parameter lists in the appendix Customized parameter list can be configured by creating a new parameter folder Flow Tool C 4 Setting the basic parameters SITRANS FM MAG 8000 124 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 125: ...to the new folder These parameters are updated and handled as the original parameters and listed in the same order as copied to the customized parameter list Flow Tool C 4 Setting the basic parameters SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 125 ...

Страница 126: ...ww siemens com flow www siemens com flow C 5 Unit selection MAG 8000 and MAG 8000 CT are delivered with totalizer and flow rate units as ordered via the MLFB structure The standard MAG 8000 units for each region are Europe m3 for totalizer and m3 h for flow rate US Gallon for totalizer and GPM Gallon per minute for flow rate Australia ML for totalizer and ML d Mega liter day for flow rate The foll...

Страница 127: ...ettings must then be affixed to the display Protected parameters The following parameters are protected see also parameter lists in appendix MAG 8000 Standard MAG 8000 CT MAG 8000 CT additionally Parameter no Parameter name Parameter no Parameter name New password 100 Actual date and time 5 Sensor tube diameter 101 Totalizer 1 7 Meter No 102 Totalizer 2 8 Totalizer unit 201 Alarm configuration lis...

Страница 128: ...n front part of PCB behind battery see connection diagram Page 33 Note Hardware key on MAG 8000 CT For MAG 8000 CT only authorized persons with the acceptance and under direction of the local authorities are allowed to use the hardware key Changing units Note Not allowed on MAG 8000 CT This function is not allowed on MAG 8000 CT 1 Select service mode and meter version upload data from meter 2 Open...

Страница 129: ... operation Be careful when writing new parameter values as meter has no default settings to return to C 6 Output configuration Pulse output can be configured as volume pulse alarm or call up Default factory setting is with output A enabled for forward flow and output B for alarm output Output configuration in Flow Tool Select I O and Com Setup The software window shows the parameters for output co...

Страница 130: ...key operation for 10 minutes Parameter 130 defines accessible display menus with selection of one or more of the following menus Operator menu Meter info menu Service menu Data logger menu Statistic menu advanced version only Tariff menu advanced version only Disabling display of menu data will not affect operation of functions C 8 Internal data handling Meter status Meter status parameter 120 giv...

Страница 131: ...s stored as a negative value Alarm and meter status are also stored for the same period to indicate alarms that have been active or that revenue data has been influenced in the specific period The logged information has a time and date stamp The data logger never stops storing data old data is overwritten following the first in first out principle i e log 1 is the last stored information which is ...

Страница 132: ...d with the real meter consumption every 4 hours The battery limit parameter 206 is the level at which the low power alarm is activated generating an alarm or call up if configured The power status parameter 513 follows the battery symbol on the display When switching between battery power from internal and external battery packs Battery power parameter 507 must be adjusted to match the actual numb...

Страница 133: ...easurement accuracy verification like the MAG VERIFICATOR FDK 083F5060 or FDK 083F5061 The MAG 8000 uses two main physical components for communication The IrDA sensor on the MAG 8000 An IrDA interface cable D 1 Commissioning The qualification certificate is generated from PDM via the following steps 1 Enabling insulation test 2 Uploading the device data to the PC 3 Generating the qualification ce...

Страница 134: ...he Qualification Certificate The procedures for enabling the insulation test are shown below Set Insulation test enable to Yes if not already enabled Select Device Download to Device to download the change to the device Qualification certificate D 2 Enabling insulation test SITRANS FM MAG 8000 134 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 135: ... available in section Insulation test in chapter Built in functions Page 89 D 3 Uploading the device data to the pc Select Device Upload to PC PG to load all device parameters from the device to the pc Qualification certificate D 3 Uploading the device data to the pc SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 135 ...

Страница 136: ...select Device Qualification Certificate The qualification certificate tool is protected by a service password that is only available for service specialists Qualification certificate D 4 Generating the qualification certificate SITRANS FM MAG 8000 136 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 137: ...Generate qualification certificate in the next dialog box A progress bar shows the percentage of the process Qualification certificate D 4 Generating the qualification certificate SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 137 ...

Страница 138: ...lues to evaluate if the device status is of proper quality or not Reference values are available in the document Qualification Certificate Reference Guideline https support industry siemens com cs ww en view 109750126 Qualification certificate D 5 Result evaluation SITRANS FM MAG 8000 138 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 139: ... 289810743 Flow rate unit parameter 9 Correction factor parameter 301 Default 1 m3 s m3 min m3 minute 60 m3 h m3 hour 3600 m3 d m3 day 86400 GPS Gallon second 264 1721 GPM Gallon minute 15850 32 GPH Gallon hour 951019 4 GPD Gallon day 22824465 MGPD 1000000 Gallon day 22 824465 CFS ft3 second 35 31467 CFM ft3 minute 2118 882 CFH ft3 hour 127132 8 L s liter second 1000 L min liter minute 60000 L h l...

Страница 140: ...Unit conversion table SITRANS FM MAG 8000 140 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 141: ...ta Logger menu L Statistic menu St Revenue menu R Factory settings marked in italic cannot be edited Note Change of units If the unit and factor is changed all unit related values become invalid It is recommended to reset all unit related values F 1 1 99 Param eter ID Meter version Display view Parameter data type Factory settings Data range Fixed parameter or meter data that are not changeable 1 ...

Страница 142: ...time PS3 production date and time year month day T hours minutes seconds 101 All O1 Totalizer 1 0 0 to 2x109 102 All O2 Totalizer 2 0 0 to 2x109 103 All O5 Totalizer 3 0 0 to 2x109 106 All Flow rate 0 to 1 25 Qn Q3 107 All Actual velocity Measured value 15 m s 108 All Flowrate percent value 0 to 125 Q4 120 All Actual flow meter status 0 to 255 binary presented with information 1 for bit 0 1 Totali...

Страница 143: ...tory settings Data range Fixed parameter or meter data that are not changeable 200 All O4 Fault status No faults 0 to 8191 binary presented with information 1 for bit 0 1 Insulation error 2 Coil current error 3 Amplifier overload 4 Database checksum error 5 Low power warning 6 Flow overflow warning 7 Pulse A overload warning 9 Consumption interval warning 10 L Leakage warning 11 E Empty pipe warni...

Страница 144: ...ommand execution 209 All Reset consumption log fault No Yes No Value is reset to No when up loading device parameters to pc after command execution 210 All Insulation alarm output enable No Yes No 211 All Insulation fault hours 0 212 All Insulation fault counter 0 213 All Insulation fault appears 2000 01 01 T 00 00 00 214 All Insulation fault disappears 2000 01 01 T 00 00 00 215 All Coil current a...

Страница 145: ...ter 0 243 All Pulse A overload fault appears 2000 01 01 T 00 00 00 244 All Pulse A overload fault disap pears 2000 01 01 T 00 00 00 250 All Consumption alarm output en able No Yes No 251 All Consumption fault hours 0 252 All Consumption fault counter 0 253 All Consumption fault appears 2000 01 01 T 00 00 00 254 All Consumption fault disappears 2000 01 01 T 00 00 00 255 All Leakage alarm output ena...

Страница 146: ...ings Data range Fixed parameter or meter data that are not changeable 300 All Totalizer volume unit factor MLFB dependent 0 to 1 x010 301 All Flow unit factor MLFB dependent 0 to 1 x010 302 All Pipe size Sensor related 25 to 1200 303 All Meter excitation frequency in battery power mode 1 15 Hz 1 60 Hz 1 30 Hz 1 15 Hz 1 5 Hz 1 5625 Hz 3 125 Hz 6 25 Hz 304 All Mains frequency MLFB dependent 50 or 60...

Страница 147: ...G 8000 Standard 7ME6810 0 05 for FW 3 07 and lower For FW 3 09 see table F 1 below 0 to 9 9 MAG 8000 CT 7ME6820 0 25 for FW 3 07 and lower For FW 3 09 see table F 2 below 15 mm s for FW 3 11 on wards 0 to 999 mm s 329 All Filter time constant 5 Tau 1 to 1000 331 All Excitation frequency limit 6 25 Hz for advanced version and 1 15 Hz for basic version 332 All Excitation frequency sensor limit Senso...

Страница 148: ...50 2 0 17 0 66 0 66 0 66 0 27 0 27 0 27 0 42 0 17 0 17 DN 65 2 0 18 0 72 0 72 0 72 0 28 0 28 0 28 0 45 0 18 0 18 DN 80 3 0 17 0 68 0 68 0 68 0 27 0 27 0 27 0 43 0 17 0 17 DN 100 4 0 17 0 67 0 67 0 67 0 27 0 27 0 27 0 42 0 17 0 17 DN 125 5 0 17 0 66 0 66 0 66 0 27 0 27 0 27 0 41 0 17 0 17 DN 150 6 0 15 0 60 0 60 0 60 0 24 0 24 0 24 0 38 0 15 0 15 DN 200 5 0 17 0 68 0 68 0 68 0 27 0 27 0 27 0 42 0 1...

Страница 149: ...DN 250 10 2 65 DN 300 12 3 82 DN 350 14 5 20 DN 400 16 6 79 DN 450 18 8 59 DN 500 20 10 60 DN 600 24 15 27 DN 700 27 20 78 DN 750 29 23 86 DN 800 31 27 14 DN 900 35 34 35 DN 1000 39 42 41 DN 1050 41 46 76 DN 1100 43 51 32 DN 1200 47 61 07 F 5 400 499 Param eter ID Meter version Display view Parameter data type Factory settings Data range Fixed parameter or meter data that are not changeable 400 Al...

Страница 150: ...All Reset communication driver No Yes No This parameter is removed from PDM parameter list The func tionality is implemented via PDM menu Device Commu nication Settings Transfer The parameter is accessible via PDM menu Device Pulse guide offline The parameter is accessible via the PDM menu Device Communication settings F 6 500 599 Param eter ID Meter version Display view Parameter data type Factor...

Страница 151: ...40 All Electrode impedance A Measured values 0 to 185 000 ohm 541 All Electrode impedance B Measured values 0 to 185 000 ohm 542 All Low medium impedance alarm 0 0 to 2 15x109 550 All Coil current disable No Yes No 551 All Fixed flow mode enable No Yes No Value is reset to No in 4 hours 552 All Fixed flow value 0 1x10x109 to 1x10x109 553 All High flow alarm limit 1 000 000 000 0 to 1x109 554 All R...

Страница 152: ...Coil current error 3 Amplifier overload 4 Database checksum error 5 Low power warning 6 Flow overflow warning 7 Pulse A overload warning 9 Consumption interval warning 10 L Leakage warning 11 E Empty pipe warning 12 C Low impedance high con ductivity warning 13 d High flow limit warning 14 15 16 Not used If none of the errors have been logged the message is No faults 614 All Latest Log period stat...

Страница 153: ...All L5 Date of log period 5 631 All L5 Log period 5 totalized 1 632 All Log period 5 totalized 2 633 All Log period 5 fault status See 613 634 All Log period 5 status informa tion See 614 635 All L6 Date of log period 6 636 All L6 Log period 6 totalized 1 637 All Log period 6 totalized 2 638 All Log period 6 fault status See 613 639 All Log period 6 status informa tion See 614 640 All L7 Date of l...

Страница 154: ...Date of log period 12 666 All L12 Log period 12 totalized 1 667 All Log period 12 totalized 2 668 All Log period 12 fault status See 613 669 All Log period 12 status informa tion See 614 670 All L13 Date of log period 13 671 All L13 Log period 13 totalized 1 672 All Log period 13 totalized 2 673 All Log period 13 fault status See 613 674 All Log period 13 status informa tion See 614 675 All L14 Da...

Страница 155: ...9 Date of log period 19 701 All L19 Log period 19 totalized 1 702 All Log period 19 totalized 2 703 All Log period 19 fault status See 613 704 All Log period 19 status informa tion See 614 705 All L20 Date of log period 20 706 All L20 Log period 20 totalized 1 707 All Log period 20 totalized 2 708 All Log period 20 fault status See 613 709 All Log period 20 status informa tion See 614 710 All L21 ...

Страница 156: ...5 status informa tion See 614 735 All L26 Date of log period 26 736 All L26 Log period 26 totalized 1 737 All Log period 26 totalized 2 738 All Log period 26 fault status See 613 739 All Log period 26 status informa tion See 614 Can be accessed via PDM menu View Log F 8 800 899 Param eter ID Meter version Display view Parameter data type Factory settings Data range Fixed parameter or meter data th...

Страница 157: ...Hz 1 5625 Hz 3 125 Hz 6 25 Hz 817 Advanced Leakage status No faults Leakage status 1 Finished successfully 2 Leakage detection running 3 Leakage detection failed Sys temStatus have fatal error 4 Leakage detection failed Empty pipe detection disabled 5 Leakage detection failed Coil current off 6 Leakage detection failed In sulation test was active during detection 7 Leakage detection stopped becaus...

Страница 158: ...ime range combination 841 Advanced R7 Date of tariff reset PS3 production date and time year month day T hours mi nutes seconds 842 Advanced Reset tariff values No Yes No Value is reset to No when up loading device parameters to pc after command execution 843 Advanced R1 Tariff1 volume 1 0 844 Advanced R1 Tariff1 period time end 360 min 6 00 24 00 0 to 1439 minutes 23 59 845 Advanced R1 Tariff1 up...

Страница 159: ...day consumption 0 869 Advanced Date of highest day con sumption PS3 production date and time year month day T hours mi nutes seconds 870 Advanced St4 Day 1 yesterday of last week consumption 0 871 Advanced Day 2 of last week con sumption 0 872 Advanced Day 3 of last week con sumption 0 873 Advanced Day 4 of last week con sumption 0 874 Advanced Day 5 of last week con sumption 0 875 Advanced Day 6 ...

Страница 160: ... Qn Q3 886 Advanced Total time in CP range 3 0 887 Advanced Upper limit in CP range 3 45 0 to 100 Qn Q3 888 Advanced Total time in CP range 4 0 889 Advanced Upper limit in CP range 4 60 0 to 100 Qn Q3 890 Advanced Total time in CP range 5 0 891 Advanced Upper limit in CP range 5 80 0 to 100 Qn Q3 892 Advanced Total time in CP range 6 0 Parameter lists F 8 800 899 SITRANS FM MAG 8000 160 Operating ...

Страница 161: ...DN 25 to 1200 1 to 48 The following table shows the relationship between flow velocity V flow quantity Q and sensor dimension DN ORZ YHORFLW Figure G 1 Sizing table SITRANS FM MAG 8000 Operating Instructions 03 2020 A5E03071515 AH 161 ...

Страница 162: ...ange 0 to 10 m s 0 to 33 ft sec Flow velocity calculation formula Metric measures V 1273 24 x Q Di2 V m s Q L s Di mm or V 353 68 x Q Di2 V m s Q m3 h Di mm Imperial measures V Q x 0 408 Pipe I D 2 V ft s Q GPM Pipe I D inch or V Q x 283 67 Pipe I D 2 V ft s Q MGD Pipe I D inch Sizing sensor G 1 Sizing table DN 25 to 1200 1 to 48 SITRANS FM MAG 8000 162 Operating Instructions 03 2020 A5E03071515 A...

Страница 163: ...Leakage detection Yes Meter utilization Yes Statistics Yes Tariff Yes Settle date Revenue Yes Features Application identification parameters 1 and 2 Tag number visible on display if numbers are selected and meter location up to 15 characters per information Time and date parameter 100 Real time clock and date max 15 minutes drift per year Totalizer parameters 101 102 and 103 3 totalizers Forward r...

Страница 164: ... to DN 600 8 to 24 1 30 Hz DN 700 to DN 1200 28 to 48 1 60 Hz Excitation frequency with mains power follows maximum sensor excitation frequency Filter constant as numbers of excitations Low flow cut off of Qn Q3 Empty pipe detection active symbol on display when active Filter selection for mains power frequency 50 60 Hz Correction factor for change of flow direction or to adjust flow measurement D...

Страница 165: ... Qn Q3 Q4 Pulse overflow on output A and B Selected pulse volume is too small compared to actual flow rate and max output pulse rate Consumption Saved data logger consumption exceeds customer selected limit on high or low consumption Leakage Leakage detected based on customer settings Advanced version only Empty pipe No water in the pipe sensor Low impedance Measured electrode impedance below cust...

Страница 166: ...ion signal chain for correct scaling Number of sensor measurements excitations Transmitter temperature battery capacity calculation Low impedance alarm for change in media Flow alarm when defined high flow exceeds Insulation test Verification mode for fast measure performance check Advanced version includes Leakage detection Consumption profile Flow statistic Insulation test Test of signal immunit...

Страница 167: ...e latest two totalized 1 index values Settling values are visible on the display Statistic Advanced version only The statistic feature delivers water consumption information to inspect the consumption limits of the installation The real values are first available after the period is finished A reset function clears the values and new values are ready after next period Until the new measured values...

Страница 168: ...Features SITRANS FM MAG 8000 168 Operating Instructions 03 2020 A5E03071515 AH ...

Страница 169: ...s area Laws and directives 11 History 8 Hotline Refer to Support request I Installation Compact 17 Indoor outdoor 17 Remote 17 Items supplied 7 L Laws and directives Disassembly 11 Personell 11 Lithium batteries 83 Return 84 Safety 12 M Mains supply 34 Maintenance 77 Manuals 115 P Parameter lists 141 PDM Commissioning 43 Pressure specifications 17 Product name 15 R Repair 77 Return procedure 83 S ...

Страница 170: ...6 System components 13 T Technical support 116 partner 116 personal contact 116 Temperature specifications 17 Test certificates 11 Type plate 15 V Vibrations 23 W Warranty 9 Index SITRANS FM MAG 8000 170 Operating Instructions 03 2020 A5E03071515 AH ...

Отзывы: