7.2
Diagnostics (menu 2)
Device status (menu 2.1)
The test routines are systematically executed during normal operation. The error is indicated by
a blinking letter on display:
D: Device error
F: Process error
S: Simulation is active
The error can also be signaled on the analog and digital outputs.
The error is shown in menu 2.1 Device status. The error messages and their assignment to the
blinking letter, the analog output and the digital outputs are listed in the table.
Table 7-1
Error messages
Error message
Blinking letter on
the display
Error message on
analog output
digital outputs 1 and
2 (relay)
Measuring path error
F
✓
3)
✓
Unreliable flow value
F
✓
3)
✓
P/F too high
F
✓
3)
✓
US gain too high
F
✓
3)
✓
Simulation active
S
-
-
COM module failure
1)
D
-
-
Measurement module failure
D
✓
✓
EEPROM failure
D
✓
✓
RAM failure
D
✓
✓
SSC failure
2)
D
-
✓
Firmware failure
D
✓
✓
1)
HART module
2)
Internal serial interface
3)
Failure signal is not output if error is suppressed and simulation function is active
Suppress error (menu 2.2)
In this menu item the error(s) "Unreliab. Flow" and/or "Meas path err." can be suppressed so that
the suppressed error(s) are not shown, neither on the display nor at the output. After every reset
or power off of the device, all error messages are available again, thus this menu setting is not
stored.
Device test (menu 2.3)
The following test sub routines are available:
Functions
7.2 Diagnostics (menu 2)
SITRANS FUS060 with PROFIBUS PA
Operating Instructions, 09/2021, A5E02225054-AD