code
Device function,
Parameters
Description
Factory setting
Setting options
Read
or write
3.6.1
Total.net unit
Physical volume unit or mass unit
Note: Whenever using mass flow the
actual density have to be selected in
menu 3.1.4
m³
l, hl, m³, Ml, ft³, Gal,
MGal, ImpGal, MIm‐
pGal, g, kg, Ton, lb
Write
3.6.2
Lo Alarm Limit
Value for lower alarm limit
-1.000.000.000 m³
-1.000.000.000 o 0
Write
3.6.3
Hi Alarm Limit
Value for upper alarm limit
+1.000.000.000 m³
-0 to 1.000.000.000 Write
3.6.4
Hysteresis
Hysteresis for the limits
0
0 to 3
Write
3.6.5
Settings net
Volume totalizer reset to “0” and
“stop/start”
No action
No action
Reset + stop
Reset + start
Write
D.1.4
Menu 4 - Device outputs
Menu
code
Device function,
Parameters
Description
Factory setting
Setting options
Read
or write
4.1
PROFIBUS
4.1.1
Bus Address
PROFIBUS address of device
126
0 to 126
Write
4.1.2
Ident-Nr
Selection of profile-specific or manu‐
fact.-specific identification number
(GSD-files: manuf. 0x8159.gsd; pro‐
file 0x9741.gsd).
Manuf. spec.
Manuf. spec.
Profile spec.
Write
4.1.3
min Tsdr
Minimum station delay time of res‐
ponder (unit: bit time =32 μs = time
for 1 bit at 31.25 kbit/s).
11 bits
Factory set (11 bits)
Read only
Read
4.2
Digital output 1
Frequency or impulse signal for
flow or alarm signal
4.2.1
Function
Assignment of output function
Pulse
Pulse
Frequency
Alarm
Flow dir. forward
Flow min
Flow max
Flow min/max
Total forw. max
Total rev. min
Total net. min
Total net max
No function
Write
4.2.2
Signal type
Configure output:
Signal: passive only
Logic: positive or negative
Passive-pos.
Passive-pos
Passive-neg
Write
HMI menu structure
D.1 Menu items (FW Rev. 4.00.00)
SITRANS FUS060 with PROFIBUS PA
Operating Instructions, 09/2021, A5E02225054-AD
111