BUS addresses
Make sure the PROFIBUS device addresses are set before operating two or more field devices on
the bus. Every address may only be assigned once so that there is no confusion. Basically, an
address range of 1 to 125 is possible; address 126 is set in the as-delivered state. Usually, the
lower addresses are assigned to the masters in PROFIBUS systems. Therefore, we recommend
starting the address for the device at 30. The addresses are set with the operation and
monitoring module in menu 4.1.1 or a cyclically use of PROFIBUS.
Note
Change bus address
The BUS address can only be changed if there is no cyclical communication and all acyclic
connections are closed.
6.3.3
Navigating the menu
The device functions and parameters available are shown in the second line of the display.
• Selection is made with the and keys. These are shown in the second line of the display.
• Enter device function or setting level of parameters with the key (Enter function).
• Exit the selected function or setting level without storage of the change by using the key
until the cursor key moves to the far left position.
Changing parameter settings
The currently valid setting appears after selecting the setting level of a parameter. If
programming is enabled, the programmable value flashes in the second line of the display. The
parameter setting can be changed. There are two different types of data input:
• Direct numerical input
• Input from given table
In the numerical input, the and keys function as cursor control. The selected digit flashes.
The key increases the flashing digit (for example 9 9 . 0 0 0 becomes 1 0 0 . 0 0 0 ).
The key decreases the flashing digit (for example 1 0 . 0 0 0 becomes 9 . 0 0 0 ).
The decimal point is moved to the right using the key and to the left using the key.
After selecting the last digit, the input is confirmed by using the key. The entered value is
accepted if it is within the permissible input range. In this case the user guidance returns to the
selection menu for the parameters of the group concerned. If the entered value is rejected, an
error message briefly appears on the display followed by the previous setting. The value can be
changed again.
Commissioning
6.3 Commissioning via local user interface
SITRANS FUS060 with PROFIBUS PA
Operating Instructions, 09/2021, A5E02225054-AD
43