Device function,
Parameters
Description
Factory setting
Setting options
Read
or write
3.1.2
Max. vol. flow
Full scale value
Note: Corresponds to 20 mA and se‐
lected max. frequency
Depends on sensor
size
See dimension table
in sensor manual;
max. 12 m/s
Write
3.1.3
Limits
Flow unit according to menu 3.1.1
3.1.3.1 Lo alarm limit
Lower alarm limit (lower limit must
be lower than upper limit)
Depends on sensor
size
See dimension table
in sensor manual
Write
3.1.3.2 Hi alarm limit
High alarm limit (higher limit must
be higher than lower limit)
Depends on sensor
size
See dimension table
in sensor manual
Write
3.1.3.3 Hysteresis
Hysteresis for limit in % of full scale
value
1%
0% to 20% of Max.
vol. flow setting
(menu 3.1.2)
Write
3.1.4.
Density
Units
Density of media
Density unit
kg/m³
g/cm³
g/l
kg/l
kg/m³
lb/ft³
lb/in³
lb/gal
lb/lmpgal
Write
Density
Input of density value for arithmetic
calculation of mass flow
+1000.00 kg/m
3
200 to 5000 kg/m
3
Write
3.1.5
Direction
3.1.5.1
Direction
Determination of main direction of
flow (forward direction)
+ Direction
+ Direction
- Direction
Write
3.1.5.2
Measurement
Choice of measurement direction
Forward only
Forward only
F reverse
Write
3.1.5.3
Hysteresis
Setting hysteresis for flow direction
related to full scale value
1%
0 to 20%
Write
3.1.6
Low flow cut
Determination of switching point for
low flow cut-off. Suppression is rela‐
ted to full scale value
1%
0 to 20%
Write
3.1.7
Flow damping
3.1.8
Time constant
Selection of time constants for meas‐
uring variables
5.0 sec
0.0 to 200 sec
Write
3.2
Sound velocity
3.2.1
Dynamic range
Unit for sound velocity
m/s
m/s
Read
3.2.2
Dynamic range
3.2.2.1
Lower value
Lower range value for sound velocity
Lower range value
< Upper range value
+600 m/s
+200 to 2000 m/s
Write
3.2.3
Limits
3.2.3.1
Lo alarm limit
Lower alarm limit
Lower alarm limit
< Upper alarm limit
+200 m/s
200 to 2000 m/s
Write
HMI menu structure
D.1 Menu items (FW Rev. 4.00.00)
SITRANS FUS060 with PROFIBUS PA
Operating Instructions, 09/2021, A5E02225054-AD
109