Operation
5.4 Operating functions
OXYMAT 64
88
Operating Instructions, 06/2008, A5E00880383-03
5.4.6.6
Reset (function 74)
74 Reset
O
2
Trigger reset
Figure 5-52 Carrying out a reset
With this function, you can carry out a warm restart.
After calling this function, the device switches to the "Warm-up phase", which activates the
functional check at the same time. The device is only completely ready for operation again
after successfully running through this phase.
5.4.6.7
Save, load data (function 75)
75 Save data O
2
Save user data
Load user data
Load factory settings
Figure 5-53 User data memory
You can configure the user data memory with this function:
Press the softkey 1
"Save user data" e.g. after the system has been successfully
commissioned. All individual settings are then saved.
Press the softkey 2
"Load user data", to load the last user data saved.
These functions are important when the device is undergoing repairs or maintenance work,
or, for example, if new parameter settings are to be tried out.
Press softkey 3 to re-establish the factory settings.
A confirmation query is set up in this function. In order to actually load the respective data in
the memory, you must confirm with "yes". If you select "no", this is cancelled.