Operation
5.4 Operating functions
OXYMAT 64
Operating Instructions, 06/2008, A5E00880383-03
57
See also
Setting the setpoints (function 22) (Page 57)
Total / single calibration (function 23) (Page 57)
Define measuring ranges (function 41) (Page 70)
5.4.3.4
Setting the setpoints (function 22)
22 Setpoints total O
2
Setpoint for zero.
: 5.4O: vpm
Setpoint for MR 1
8.5O vpm
Setpoint for MR 2
12.5O vpm
Setpoint for MR 3
6OO.5O vpm
Setpoint for MR 4
8245.5O vpm
Figure 5-12 Setting the setpoints
The function screen shows the setpoint input with total calibration. The third measuring
range is chosen here as the leading measuring range.
With single calibration, there is no choice for the leading measuring range.
The definition of the leading measuring range does not apply to Autocal mode.
5.4.3.5
Total / single calibration (function 23)
23 Calibr. settings O
2
Total range calibration
Figure 5-13 Select calibration method
With this function, select between the total and single calibration of the measuring ranges.
Total calibration means that you adjust one "leading measuring range" and all other
measuring ranges are calibrated using the switching ratio.
If this function is not activated, each measuring range is calibrated separately.