Operation
5.4 Operating functions
OXYMAT 64
78
Operating Instructions, 06/2008, A5E00880383-03
5.4.5.9
Date/time (function 58)
58 Date/Time O
2
New Date
(dd.mm.yy;24h/day)
New Time:
14 : 55
Set clock
Actual Date
Actual Time
26.O3.2OO7 14:55
Figure 5-43 Setting the date/time
With this function, you can set the date and time.
The cursor is already located at the first place of the date display. Enter the new settings in
the order: day, month, year. By pressing [ENTER], jump to the next input field. Set the time
in the same way (hours, minutes).
The system clock of the device is not buffered against mains failures (not a real-time clock).
NOTICE
When the device is switched off, the clock stops and is not updated.
The settings are especially important for troubleshooting. Errors which are always stored in
the logbook (function 60) can be allocated more easily with the help of the date and time.
Press the third softkey
"Set clock" in order to accept the set data. These then appear at the
bottom edge of the display.
See also
Logbook (function 3) (Page 53)
Logbook settings (function 60) (Page 80)