The X and Y values for the safety laser scanner can be determined by measuring the
offset between the center point of the safety laser scanner and the table corner. If you
calculate the offset of the safety laser scanner to the table corner with the coordinates
of the table corner, you get the values for
X
and
Y
.
X
Y
Offset between safety laser
scanner and the lower right
table corner
80 mm
– 80 mm
Coordinates of the lower right
table corner
1,200 mm
0 mm
Position values for safety
laser scanners
1,280 mm
-80 mm
Complementary information
Safety-relevant values are marked with the following symbol:
6.3.3.7.3
Configuring name of the safety laser scanner
Important information
NOTE
These settings are optional and not required for operation of the safety system.
Prerequisites
•
You are registered as a service technician.
Approach
1.
Click on the desired safety laser scanner. Placeholders with assigned safety laser
scanner are orange.
2.
Click on
Name
.
3.
In the
Name
field, enter the desired name.
4.
Click on
Apply
.
✓
The
Success
dialog is displayed. Confirm dialog to continue.
6.3.3.7.4
Configuring the connection data
Prerequisites
•
You are registered as a service technician.
Approach
1.
Click on the desired safety laser scanner. Placeholders with assigned safety laser
scanner are orange.
2.
Click on the
Address
tab.
3.
In the
IP Address
,
Subnet
and
Router
fields, enter connection data.
4.
Click on
Apply
.
✓
The
Success
dialog is displayed. Confirm dialog to continue.