TROUBLESHOOTING
6
6.2
Sender/receiver unit
Malfunctions
Warning and error messages in SOPAS ET
Fig. 59: SOPAS ET menu: DH SP100/Diagnosis/Error messages/Warnings
Current warning or error messages, or earlier messages stored in the error memory, can be
shown by selecting “actual” or “memory” in the “Error selection” or “Warnings selection”
window.
●
Display of error or warning: With LED symbol
●
Description of error or warning: In the description field of SOPAS ET
Malfunctions listed below can probably be cleared onsite.
Symptom
Possible cause
Action
●
LEDs of the
sender/receiver
are not on
●
No laser beam
●
No supply voltage
●
Connection line not connected correctly
or defective
●
Defective plug connector
▸
Check plug connectors and lines.
▸
Contact SICK Customer Service.
Message
Significance
Possible cause
Action
Contamina-
tion
Current reception intensity
is lower than the allowable
limit value (
●
Deposits on the optical
interfaces
●
Unclean purge air
▸
Clean the optical surfaces (
on the sender/receiver unit”, page 84
▸
Check purge air filter (
the purge air supply”, page 88
▸
Contact SICK Customer Service.
Span test,
Zero point
Deviation from setpoint
value > ±2%.
Sudden changes in measur-
ing conditions during deter-
mination of control values
▸
Repeat the function check.
▸
Contact SICK Customer Service.