START-UP AND PARAMETER SETTINGS
4
Defining regression coefficients
The “Calibration coefficients for calculation of concentration with scattered light” field
serves to select a coefficients set freely (set 0) or a fixed default setting (sets 1 to 3).
Select set 0 when the measuring system must be calibrated (
see “Calibration for dust con-
centration measurement”, page 69
Sets 1 to 3 can be used when the highest precision is not decisive.
Set
Regression coefficients
setting
Typical application
Regression coefficients
Quadratic Linear
Absolute
Set 0
Freely selectable
Any
0
1
0
Set 1
Fixed: For small particle
sizes (on average 2
μ
m)
Usage after cloth filter
plants
0
0.5
0
Set 2
Fixed: For medium particle
sizes (on average 5
μ
m)
Usage after electrofilter
0
2
0
Set 3
Fixed; for large particle sizes
(on average 10
μ
m)
Usage after coarse filter
(cyclone filter)
0
3
0
The regression coefficients in sets 1 to 3 refer to dusts with an average density
of 2.5 g/cm³, almost spherical particle structure and even dust distribution across the
duct cross-section.