DATOSPIR TOUCH User’s Manual 31
511-B00-MU2 • REV. 1 .09
3
Discharge the syringe once at every one of the following flow
rates between 0-12L/s:
Low flow level: 0,4 – 1,2 L/s
Mid-flow level: 2 - 5 L/s
High flow level: 6 - 12 L/s
To simplify the procedure, the spirometer allows you to carry out a
single maneuver at Mid-flow level: 2 - 5 l / s.
4
Press
to cancel the maneuver or
to accept it and go to
the next range of flow.
5
At the end of the operation, press
to see the results of the
maneuvers performed: the percentages of variation and
calibration factors calculated as the average of all the maneuvers
performed on a range flow.