DATOSPIR TOUCH User’s Manual 61
511-B00-MU2 • REV. 1 .09
Interpretation NLHEP. (Only valid for predicted values that
calculates the LLN. For example: Hankinson). Ferguson et al.
Office Spirometry for Lung Health Assessment in Adults.
Chest 2000; 117: 1146-1161.
6.6 TRANSDUCERS
RANGES AND MEASUREMENTS (According to ATS/ERS2005)
Fleisch
Turbine
Lilly
Dispos.
Measurement Scale (BTPS)
0 to ±16 l/s
0 to 10 l
Flow
Volume
Dynamic flow resistance
< 1,47 hPa (1.5 cmH
2
0) / (l/s) a 14 l/s
Precision of measurements (BPTS)
(the highest value)
3% or 50 ml
5% or 200 ml/s
10% or 300 ml/s
Volume
Flow
PEF
Time related precision
0,50%
Volume resolution
<10ml
<10ml
<10ml
Sampling frequency
100Hz
100Hz
100Hz
Transducer lifetime
1400
disinfect.
or 3 years
1400
disinfect.
or 3 years
Only one
use (Expir.
3 years)
6.7 MANUFACTURER’S LIABILITY
SIBEL S.A.U. is only liable for the safety, reliability and working
order of this equipment if:
The place where the equipment is installed or used is compliant
with the requirements related to the IEC electrical installation
and other applicable regulations.
All repairs, services and modifications inside and outside the
guarantee period are carried out by SIBEL S.A.U. technical staff.
The equipment is used by skilled staff according to the
recommendations of this User’s Manual.