DATOSPIR TOUCH User’s Manual 65
511-B00-MU2 • REV. 1 .09
2
A second procedure, which can be performed by the user,
consists of regular monitoring of the appearance of the different
connections and other external parts of the equipment. Check
that all connections are perfectly connected, that no cable and/or
connector or any other element is broken or damaged.
3
Execute a calibration check with a 3-L syringe. Calibration
checks must be undertaken daily.
4
Define the periods in days between calibrations or preventive
maintenance work on the
CUSTOMIZATION MENU
of the
equipment. If the days specified are exceeded, the equipment
warns of such by displaying a sign every time it is started. If “0”
days is entered, a warning is never given.
In the event of detecting any problem that the user cannot solve,
contact the SIBEL S.A.U. After-Sales Service or your
distributor to review or repair it.
Actions carried out by skilled technical personnel:
According to the different regulations, particularly the 93/42/EEC
Medical Device Directive, electromedical devices should be
verified and/or calibrated regularly to ensure reliable functions and
the safety of patients, users and the environment.
This Technical check will be performed every year following
the DATOSPIR TOUCH Verification and Adjustment Procedures,
available from the manufacturer SIBEL, S.A.U. This type of
operation must be carried out by skilled technical staff from the
distributor’s or manufacturer’s technical service.
On all accounts, SIBEL S.A.U., as the manufacturer, must
provide written authorization, for at least the guarantee period, for
the corresponding technical personnel to carry out said
maintenance and will not be held liable under any circumstances
for any damage, malfunction, etc. that may arise as a result of
defective maintenance by people not employed by SIBEL S.A.U.