Guidance and manufacturer’s declaration – electromagnetic immunity
DATOSPIR TOUCH is intended for use in the electromagnetic environment specified below. The costumer or
the user of DATOSPIR TOUCH should assure that it is used in such an environment.
Below mentioned tests are applicable to enclosure. Additionally, test EN-IEC 61000-4-2 is applicable to
input/output signal and patient connections.
Immunity test
Immunity test
Immunity test
Immunity test
Electrostatic
discharge (ESD)
EN-IEC 61000-4-2
±8 kV contact
±2, ±4, ±8, ±15 kV, air
±8 kV contact
±2, ±4, ±8,
±15 kV, air
Radiated RF
EN-IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
80% AM at 1kHz
3 V/m
Proximity fields from
RF communications
EN-IEC 61000-4-3
(View following table).
Portable and mobile RF communications
equipment should be used no closer than
30cm to any part of DATOSPIR TOUCH,
including cables. The distance is
calculated according to:
where
P
is the output power rating of the
transmitter in watts (W), d is the
recommended separation in meters (m)
and E is the immunity level in V/m.
Power frequency
(50 / 60 Hz)
magnetic field
EN-IEC 61000-4-8
30 A/m
30 A/m
Only applicable to devices sensitive to
magnetic fields.