41
7.6 EVT4 Action Group
EVT4 Action group is indicated only when EVT3, EVT4 Outputs (EVT3 option) are/is
ordered. To enter the EVT4 Action group, follow the procedure below.
1
Press the
key 5 times in ORP Display mode or Cleansing Output
mode.
2
Press the
key.
The unit proceeds to the EVT4 Action group, and ”EVT4 type“ will appear.
Character
Setting Item, Function, Setting Range
Factory Default
EVT4 type
No action
• Selects an EVT4 output (Contact output 4) type. (Fig 7.6-1, p.46)
Note: If EVT4 type is changed, EVT4 value defaults to 0.
•
: No action
: ORP input low limit action
: ORP input high limit action
: Cleansing output
: ORP input error alarm output
EVT4 value
0 mV
• Sets EVT4 value. (Fig 7.6-1, p.46)
• Available when
(ORP input low limit action) or
(ORP input high limit action) is selected in [EVT4 type].
• Setting range: Input low limit to Input high limit
EVT4 proportional band
0 mV
• Sets EVT4 proportional band. (Fig 7.6-1, p.46)
ON/OFF control action when set to 0.
• Available when
(ORP input low limit action) or
(ORP input high limit action) is selected in [EVT4 type].
• Setting range: 0 to 4000 mV
EVT4 reset
0 mV
• Sets EVT4 reset value.
• Available when
(ORP input low limit action) or
(ORP input high limit action) is selected in [EVT4 type].
• Not available for the ON/OFF control action.
• Setting range:
200 mV