EinScan HX User Manual
18
3.1.2 Standard Calibration
In the initial use, you will be guided to
Standard Calibration
directly.
Steps
1.
Place the position paper on a horizontal flat surface.
Figure 3-1 Position Paper
2.
Open and then put the calibration board vertically aligning to the black line frame
Figure 3-2 Calibration Board
3.
Hold the scanner face to the center of board in upright position. Move the mark
into the
white line frame.
Figure 3-3 Preview of the Calibration Board
4.
Press the play button on the scanner. Keep adjusting the distance between the scanner and the
calibration board by moving the scanner horizontally from near to far slowly until all the color
bars turn green, which means that all the required distances have been captured.