RB-473-C02G
30
4)
Adjustment of AXIS
Input the value of the test lens using and
.
Place and align the cylindrical lens of +5D so that
AXIS value becomes 90
.
When the prisms reach between 0.00 and 0.01 in
both X and Y direction, CYL and AXIS are
automatically calibrated to the value of the test lens.
Input the value of the test lens using and
.
Place and align the cylindrical lens of +5D so that
AXIS value becomes 180
.
When the prisms reach between 0.00 and 0.01 in
both X and Y direction, CYL and AXIS are
automatically calibrated to the value of the test lens.
5)
Prism adjustment
・
Adjustment of 2
△
and 5
△
Align the prism lens so that the base direction
becomes 0°.
The prism value is automatically calibrated to the
value of the calibration lens.
Use
and
to input the calibration
value.
・
Adjustment of 10
△
Align the calibration lens according to the indication
of the base direction (Align B).
Base direction the measured value is taken in:
0°, 45°, 90°, 135°, 180°, 225°, 270°, 315°
When the measured value of the indicated base
direction is taken in, a beep sounds and the
indication of the base direction changes.
When the measured value of the last alignment
(base direction: 315°) is taken in, the beep sounds
and calibration is automatically done.
Calibration
S
C
A
P
X
P
Y
0.00
5.00
90
O 0.00
D 0.00
12:+5D A 90
+5.00D
Place the C +5D lens.
Marking OK
Abbe 60
CYL +
Step 0.01
Auto
Naked
Prism value
Calibration
S
C
A
P
B
0.00
0.00
180
5.00
0
15:P 5△
5.00△
Place the P 5△ lens.
Marking OK
Abbe 60
CYL ‑
Step 0.01
Auto
Naked
Calibration
S
C
A
P
X
P
Y
0.00
5.00
90
O 0.00
D 0.00
13:+5D A180
+5.00D
Place the C +5D lens.
Marking OK
Abbe 60
CYL +
Step 0.01
Auto
Naked
Prism value
Base direction
Calibration
S
C
A
P
B
0.00
0.00
180
10.00
0
16:P10△
10.00△
Align B 0
Marking OK
Abbe 60
CYL ‑
Step 0.01
Auto
Naked
Base direction indication
of alignment