RB-473-C02G
29
1)
Start of calibration
Follow the instructions on the screen.
The screen shown on the right is the start screen of '6.
Calibration' (or '5.Calibration'). Press switch to
call up the next screen.
2)
0D calibration
In about two seconds, all data are automatically
calibrated to 0D.
3)
Calibration of spherical lenses (-5D to +25D)
Input the value of the test lens using and
.
Place and align the corresponding test lens so that
each of prism X direction and Y direction becomes
0.00.
When the prism value reach between 0.00 and 0.01
in both X and Y directions, the spherical value is
automatically calibrated to the value of the test lens.
Calibration
S
C
A
P
X
P
Y
0.00
0.00
180
O 0.00
D 0.00
00:Start
Take the lens away and push −> key.
Abbe 60
CYL ‑
Step 0.01
Calibration
S
C
A
P
X
P
Y
0.00
0.00
180
O 0.00
D 0.00
01:0D
0 . 0 0 D
Take the lens away and push −> key.
Abbe 60
CYL ‑
Step 0.01
For spherical lenses, perform calibration to No. 02 to No. 05 only though six
more items (No. 06 to No. 11) are included in the parameter list Skip No.06 to
No. 11 with
switch.
Calibration
S
C
A
P
X
P
Y
‑5.00
0.00
180
O 0.00
D 0.00
02:‑5D
‑5.01D
Place the S ‑5D lens.
Marking OK
Abbe 60
CYL ‑
Step 0.01
Auto
Naked
Value of test lens
Prism value