RB-473-C02G
23
7.3. Image samples
Normal
No missing parts in the measurement ring shown.
Nothing is shown in the background (all black).
The measurement values are stable.
Abnormal
When unnecessary light exists:
The background is uneven.
The measurement values are unstable.
'Retry Error' may appear.
When the lens being measured is extremely dirty or has
some flaws, and /or the lens under the lens stand is dirty:
A part of the measurement ring is missing.
The measurement values are unstable.
When the missing part is large (75% or more of the
whole ring), 'Retry Error' is indicated.
Note: SLM-4000 uses a black-white LCD with low resolution and the ring may seem chipped. When the ring
misses some part, check as follows.