RB-473-C02G
28
8.3. How to engage in Calibration screen
1)
Select '6. Calibration' (or '5. Calibration') with or switch. Press switch, and the calibration
screen is indicated.
2)
Press switch to escape from the calibration screen.
8.4. Calibration procedure
Calibration parameter list
No.
Item
Setting value
Description
00
Start
Start
01
0D
0.00D
0D calibration
02
S
–
5D
- 5.000D
03
S
–
10D
-10.000D
04
S
–
15D
-15.000D
05
S
–
20D
-20.000D
06
S
–
25D
-25.000D
07
S +5D
+ 5.000D
08
S +10D
+10.000D
09
S +15D
+15.000D
10
S +20D
+20.000D
11
S +25D
+25.000D
Calibration of spherical values at each setting value
(automatic setting)
12
C +5D A90
+ 5.00DD
13
C +5D A180
+ 5.00D
Overall calibration of spherical and cylindrical
values and calibration of angle to the lens plate.
14
P 2
△
+ 2.00
△
Calibration of 2 prism powers.
15
P 5
△
+ 5.00
△
Calibration of 5 prism powers.
16
P 10
△
+10.00
△
Calibration of 10 prism powers.
17
Far point
0.12 (standard value)
Progressive lens: calibration of far point.
18
Near point
0.15 (standard value)
Progressive lens: calibration of near point.
19
Auto judge
0.18 (standard value)
Judgment value for progressive lens
20
Finished
End
For
part, the factory setting value is valid. If you change it, you may not be able to
measure correctly.
In '6.Calibration' (or '5.Calibration'), each setting and taking in the calibration data are automatically done in order
to facilitate the calibration procedure.