1 OUTLINE
MUX-100 SERVICE MANUAL 1 - 3
MUX-100D
MUX-100DJ
High-voltage generator
Nominal electric power
12.5kW (100kV
、
125mA
、
0.1s)
(
The product of the current allowed to flow for 0.1 seconds at a tube
voltage of 100 kV and tube voltage
)
The nominal maximum tube voltage and
the maximum tube current allowed to flow
at the nominal maximum tube voltage
125kV
、
100mA
The maximum tube current and the
maximum tube
voltage that allows the maximum tube
current to flow
160mA
、
80kV
The combination of the tube voltage and
tube current that output the maximum
electrical power
125kV
、
100mA
100kV
、
125mA
80kV
、
160mA
The minimum value of the current time
product
0.32mAs
The maximum value of the current time
product
320mAs 320mAs
Ratings
The minimum value of nominal irradiation
time
3.2ms
40
~
90kV 0.32
~
320mAs 0.32
~
320mAs
91
~
100kV 0.32
~
280mAs 0.32
~
280mAs
101
~
110kV 0.32
~
250mAs 0.32
~
250mAs
111
~
120kV 0.32
~
220mAs 0.32
~
220mAs
The current time product
that can be set with the
X-ray tube voltage
121
~
125kV 0.32
~
200mAs 0.32
~
200mAs
Range of the setting:
40
~
125kV
、
in 1-kV increments
Setting and display of
tube voltage for
exposure
Display: Digital
display
Accuracy X-ray
tube voltage
Less than 10%
Range of the setting:
0.32
,
0.36
,
0.40
,
0.45
,
0.50
,
0.56
,
0.63
,
0.71
,
0.80
,
0.90
,
1.0
,
1.1
,
1.2
,
1.4
,
1.6
,
1.8
,
2.0
,
2.2
,
2.5
,
2.8
,
3.2
,
3.6
,
4.0
,
4.5
,
5.0
,
5.6
,
6.3
,
7.1
,
8.0
,
9.0
,
10
,
11
,
12
,
14
,
16
,
18
,
20
,
22
,
25
,
28
,
32
,
36
,
40
,
45
,
50
,
56
,
63
,
71
,
80
,
90
,
100
,
110
,
125
,
140, 160
,
180
,
200
,
220
,
250
,
280
,
320mAs
The maximum value is 200 mAs with MUX-100HJE.
Setting and display of
radiographic current time
product.
Display: Digital
display