6 ADJUSTMENT/REPLACEMENT OF EACH PART
MUX-100 SERVICE MANUAL
6 - 3
6-1-3
How to remove the Top Cover
1.
Turn the X-ray tube 180 degree around the X-ray tube stand. And move the X-ray
tube to the lowest position.
Turn off the key switch and the main breaker.
2.
Unscrews the three screws as shown in Fig. 6-4.
3.
Unscrews the three screws on the Top cover
4.
Pull up the front part of the top cover (Fig. 6-5).
Fig. 6-4 Screw clamp positions of the top cover.
Fig. 6-5 Screw clamp positions of the top cover.
CAUTION
The top cover and the main unit are connected with the cables.
Be careful to remove the top cover.