4 INSTALLATION AND MAINTENANCE/INSPECTION
MUX-100 SERVICE MANUAL
4 - 16
Ⅳ
Setting of auto power OFF
At first, select the auto power OFF setting in user setting mode.
At this time, “APO” (auto power OFF setting) is displayed in the tube voltage display unit, and the current preset
value is displayed in the tube current time product display unit.
NOTE
At the time of shipment from the factory, this item is set to “15” (The power automatically
turns OFF 15 minutes later.).
Change the preset value using the tube current time product setting keys, then press the registration key to
register the contents of setting. Then, “reCord” is displayed in the display unit, the contents of setting are
registered, and the system returns to the normal use status.
Tube current time product display
(setting code)
Contents
--
The auto power OFF function is not valid.
05
The power automatically turns OFF 5 minutes later.
10
The power automatically turns OFF 10 minutes later.
15
The power automatically turns OFF 15 minutes later.
20
The power automatically turns OFF 20 minutes later.
30
The power automatically turns OFF 30 minutes later.
Table 4-8 Contents of Setting of Auto Power OFF
Fig. 4-19 How to Set the Auto Power OFF
The setting code is
displayed here.
Change the setting code
using these keys.
At the end, press and hold
the registration key until
“reCord” is displayed.