4 INSTALLATION AND MAINTENANCE/INSPECTION
MUX-100 SERVICE MANUAL
4 - 13
Ⅱ
Selection of battery type
Set the type of batteries mounted in the system.
Set the battery code displayed in the tube current time product display unit (mAs) on the operation panel using the
tube current time product setting key (
△
). (For the contents of setting, refer to Table 4-5.) When the setting is
completed, press and hold the registration key until “CALend” is displayed.
NOTE
At the time of shipment from the factory, the battery code is set to “4”. The setting does not
have to be changed in the installation work.
NOTE
Compared with the standard battery, the number of available scans, the travel distance, etc.
offered by the LC-P127R2 manufactured by Matsushita Battery Industrial by one-time charging
are considerably worse. It is recommended to use the standard battery so that the system can
offer its full performance.
CAUTION
The characteristics vary depending on the battery type. If a wrong code is set, scans, traveling
and charging may not be performed correctly. In the worst case, electrical parts may be burnt
or fire may occur.
Set code
Type of Battery (Manufacturer)
0
OLD PXL12072 (Japan Storage Battery)
Manufacture date No.-000703
1
LC-P127R2 (Matsushita Battery Industry Co. Ltd.)
2 Not
used
3
NEW PXL12072 (Japan Storage Battery)
Manufacture date No.-000712
4
PX12090 (Japan Storage Battery)
Table 4-5 Contents of Setting of Battery Type
Fig. 4-17 How to Set the Battery Type
The
setting code
is displayed here.
Change the setting code by
pressing this key.
Every time this key is pressed, the
displayed number changes in the
order (4
Æ
0
Æ
1
Æ
2
Æ
3
Æ
4)
.
At the end, press and
hold the registration
key until “CALend” is
displayed.
Fig. 4-15
Panel Display in the Initial Setting
Mode