87
Set the lens voltages with the Peak Monitor and confirm the output voltages at the
feedthrough contacts on the IS CTRL-A PCB.
The normal output voltage lies in the range: set voltage
±
5%.
LENS1, LENS2, LENS3, LENS4
If the output voltages do not lie within the set ranges, conduct the checks described in
2.
Checking the lens voltages
for MAIN CTRL-A (page 60).
3. Checking the filament currents
When a filament is on, the filament current can be determined from the following voltage
measurements. The relationship between the filament current and the voltage measured
with the probe on CP16[FIL] and the - probe on CP18[EV] is as follows:
Filament current (A) = 2 x (V
CP16
– V
CP18
) max about 3.5A
If (V
CP16
– V
CP18
) = 1.4V, the filament current is 2.8A.
A filament discontinuity is detected and an error occurs if (V
CP16
– V
CP18
)
≤
approx. 0.6V, that
is if the filament current
≤
approx. 1.2A.
The discontinuity in the filament can be easily confirmed by the following method.
Turn off the filament and measure the resistance between F1+ and F1-.
If resistance value
≤
1
Ω
, the filament is normal.
Be aware of the polarity when measuring the resistance; correct measurement is only
possible by applying a positive voltage to F1+ and a negative voltage to F1-. If the polarity
is unknown, turn off the instrument power, remove IS CTRL-A to eliminate the influence of its
circuits and measure the resistance directly at the feedthrough terminals.
4. Checking the TRAP current control circuit
Set the emission current, turn the filament on, and measure the voltages at CP12[EMC],
CP13[TRP], and CP14[+10A].
If V
CP14
≠
10V, the IS CTRL-A PCB is probably defective.
V
CP12
=set current(
µ
A)/100
If emission current = 60
µ
A, V
CP12
=0.6V.
Any abnormality suggests an incorrect D/A voltage from the CPU PCB.
V
CP12
=V
CP13
Any abnormality could be due to a filament discontinuity or incorrect installation.
5. Checking the emission voltage control circuit
This circuit generates the emission voltage (-10V to -200V).
Set the emission voltage, turn the filament on, and measure the voltages at CP15[EMV],
CP18[EV], and CP19[INV].
V
CP15
= 0.05 x set voltage (V)
If emission voltage = -70V, V
CP15
=-3.5V.
Содержание GCMS-QP2010 Series
Страница 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Страница 14: ...14 Do not increase the ion source or interface temperature above the level required for analysis...
Страница 27: ...27...
Страница 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Страница 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Страница 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Страница 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Страница 79: ...79 DC POWER A Block Diagram...
Страница 84: ...84 MAIN CTRL A Block Diagram...
Страница 85: ...85...
Страница 90: ...90 IS CTRL A Block Diagram...
Страница 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Страница 98: ...98 IG CTRL A PCB Block Diagram...
Страница 103: ...103 CP6 24V normal range 22 7 to 25 3VDC CP7 24V normal range 22 7 to 25 3VDC CP12 5V normal range 4 8 to 5 3VDC...
Страница 109: ...109 DI CTRL A block diagram...
Страница 145: ...121 5 Remove the IS CTRL A Board Unit Disconnect the two connectors Remove the three screws and the spacer screw...
Страница 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Страница 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Страница 154: ...130 10 Preparation of new TW300 New TW 300 Removed TW300 Check the name plate Old type New type...
Страница 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Страница 158: ...134 Hook the right side hooks on the neck of inlet Attach the left side screws and hooks Fasten lightly the 4 screws...
Страница 160: ...136 Remove the cardboard Completion of TMP mounting...
Страница 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Страница 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...
Страница 164: ...140 14 Place the covers Place the front cover Place the front door...