60
Problems when starting evacuation
5. No ready status after automatic start (cont.)
Check PG.
□
Does the PG value fluctuate?
The turbo molecular pump (TMP) does start above 300kPa.
If no leaks exist, the PG could be defective. See section
4.6 PCB Assy, Main
CTRL-A
.
A discontinuity in the PG wiring causes an error message to be display and prevents
startup.
Check the TMP.
□
Is the TMP running?
A problem could exist in the MAIN POWER PCB or cable if the TMP does not run
although PG is less than 300Pa.
−−
> See section
4.4 Main Power Assy
.
□
Has a TMP power supply error occurred?
The status LEDs can be seen by removing the sheet metal between the GC and MS.
If an error is displayed for TMP power supply, this indicates a problem with the TMP
or TMP power supply. Replace one of these and check the condition again.
Caution
The power supplies for the large TMP and small TMP are incompatible in early
GCMS-QP2010 models. Never swap the wiring for testing purposes or normal
operation will not be subsequently possible.
Hint
Evacuation system protective operations
(1) Start
Does the flow become READY within 3 minutes? No
−−
> error
(2) RP ON
Does the Pirani gauge drop below 300Pa within 5 minutes of turning on the rotary
pump (RP)? No
−−
> error
Содержание GCMS-QP2010 Series
Страница 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Страница 14: ...14 Do not increase the ion source or interface temperature above the level required for analysis...
Страница 27: ...27...
Страница 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Страница 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Страница 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Страница 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Страница 79: ...79 DC POWER A Block Diagram...
Страница 84: ...84 MAIN CTRL A Block Diagram...
Страница 85: ...85...
Страница 90: ...90 IS CTRL A Block Diagram...
Страница 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Страница 98: ...98 IG CTRL A PCB Block Diagram...
Страница 103: ...103 CP6 24V normal range 22 7 to 25 3VDC CP7 24V normal range 22 7 to 25 3VDC CP12 5V normal range 4 8 to 5 3VDC...
Страница 109: ...109 DI CTRL A block diagram...
Страница 145: ...121 5 Remove the IS CTRL A Board Unit Disconnect the two connectors Remove the three screws and the spacer screw...
Страница 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Страница 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Страница 154: ...130 10 Preparation of new TW300 New TW 300 Removed TW300 Check the name plate Old type New type...
Страница 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Страница 158: ...134 Hook the right side hooks on the neck of inlet Attach the left side screws and hooks Fasten lightly the 4 screws...
Страница 160: ...136 Remove the cardboard Completion of TMP mounting...
Страница 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Страница 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...
Страница 164: ...140 14 Place the covers Place the front cover Place the front door...