MAIN CTRL-A.
A variety of causes could lead to lack of signal output or small signals. If the
problem is not resolved by the checks above, check each electrical unit in the
sequence below.
Vacuum system
Low degree of vacuum can lead to poor peak strength and unfamiliar peaks may appear.
Filament In
principle,
defective control causes an error display. Check the voltage settings.
Ion source
Defective insulation, defective assembly, or defective installation can result in no peaks.
Lenses
Defective electrical system or incorrect connection can result in no peaks.
Rods, RF source Incorrect wiring or a defective electrical system can result in no peaks.
CDD
Incorrect wiring or a defective electrical system can result in almost no peaks.
EM
Incorrect wiring or a defective electrical system can result in no peaks.
Signal system
Incorrect wiring or a defective electrical system results in no response to vibrations.
Содержание GCMS-QP2010 Series
Страница 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Страница 14: ...14 Do not increase the ion source or interface temperature above the level required for analysis...
Страница 27: ...27...
Страница 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Страница 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Страница 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Страница 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Страница 79: ...79 DC POWER A Block Diagram...
Страница 84: ...84 MAIN CTRL A Block Diagram...
Страница 85: ...85...
Страница 90: ...90 IS CTRL A Block Diagram...
Страница 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Страница 98: ...98 IG CTRL A PCB Block Diagram...
Страница 103: ...103 CP6 24V normal range 22 7 to 25 3VDC CP7 24V normal range 22 7 to 25 3VDC CP12 5V normal range 4 8 to 5 3VDC...
Страница 109: ...109 DI CTRL A block diagram...
Страница 145: ...121 5 Remove the IS CTRL A Board Unit Disconnect the two connectors Remove the three screws and the spacer screw...
Страница 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Страница 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Страница 154: ...130 10 Preparation of new TW300 New TW 300 Removed TW300 Check the name plate Old type New type...
Страница 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Страница 158: ...134 Hook the right side hooks on the neck of inlet Attach the left side screws and hooks Fasten lightly the 4 screws...
Страница 160: ...136 Remove the cardboard Completion of TMP mounting...
Страница 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Страница 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...
Страница 164: ...140 14 Place the covers Place the front cover Place the front door...