36
Description of Parts
8. Turbo Molecular Pump
Two turbo molecular pumps (TMP), one for the ion-source unit exhaust (TW-300) and one
for the analyzer exhaust (TW-70), are connected to a single rotary pump (E2M1.5).
Removing the Turbo Molecular Pumps
1. Remove IS Cont PCB. Remove the IG tube.
2. Disconnect the TMP exhaust unit from the rotary pump. Loosen and remove the
clamp from the KF flange. The TMP1 flange is integral with the TMP body and can be
difficult to remove.
3. Disconnect the connector (power) from the power supply.
4. Remove the TMP fixing screws and remove the TMP.
5. Lower TMP1 to the top of the TMP power supply, slide it to the left (as seen from the
front), and remove it from the instrument.
6. Reverse this procedure to install the TMP.
Caution Ensure that no dust or dirt sticks to the O-ring at the top of the TMP.
(Purge with dry gas (He, N
2
) during installation.)
RP connection (TMP exhaust unit)
TMP fixing screws
(4 positions)
TMP1
(TW300)
TMP2
(TW70)
Содержание GCMS-QP2010 Series
Страница 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Страница 14: ...14 Do not increase the ion source or interface temperature above the level required for analysis...
Страница 27: ...27...
Страница 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Страница 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Страница 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Страница 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Страница 79: ...79 DC POWER A Block Diagram...
Страница 84: ...84 MAIN CTRL A Block Diagram...
Страница 85: ...85...
Страница 90: ...90 IS CTRL A Block Diagram...
Страница 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Страница 98: ...98 IG CTRL A PCB Block Diagram...
Страница 103: ...103 CP6 24V normal range 22 7 to 25 3VDC CP7 24V normal range 22 7 to 25 3VDC CP12 5V normal range 4 8 to 5 3VDC...
Страница 109: ...109 DI CTRL A block diagram...
Страница 145: ...121 5 Remove the IS CTRL A Board Unit Disconnect the two connectors Remove the three screws and the spacer screw...
Страница 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Страница 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Страница 154: ...130 10 Preparation of new TW300 New TW 300 Removed TW300 Check the name plate Old type New type...
Страница 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Страница 158: ...134 Hook the right side hooks on the neck of inlet Attach the left side screws and hooks Fasten lightly the 4 screws...
Страница 160: ...136 Remove the cardboard Completion of TMP mounting...
Страница 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Страница 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...
Страница 164: ...140 14 Place the covers Place the front cover Place the front door...