6
Day 1
1. Checking Installation Conditions
Confirm the installation conditions, according to the Installation Completion
Checksheet (ZEAV-0416).
Check the power supply.
1)power requirements
Frequency requirements
MS power supply: 115V
±
5%, 1000VA
230v
±
5%,1000VA
GC power supply: 115V
±
5%, 1800VA
230V
±
5%,2600VA
230V
±
5%, 2600VA for high-power oven specification.
Not including GC options.
In addition,power for the pc,monitor,and printer must be provided.
*There may be instance in which the current capacity will increase as options are
attached.
2) Ground : 100
Ω
or less
3)Voltage Fluctuation Range
Specified power supply voltage range:
±
5%
Operating power supply voltage range:
±
10%
CAUTION: USE the PC,monitor,and printer that conform to the law and regulation in
respective
NOTE: Be sure sure that power is supplied from a apower sourse with a ground fault
breaker in
NOTE: Voltage fluctuations are established so that the instrument can operate at full
capacity within the range of
±
10
%
,including any transient noise in the AC line.
Note,however,that the voltage fluctuation range that guarantees the performance
specifications should be used within
±5%
.
・
The degree of frequency stability of
power source should be within
±
0.5%.
Содержание GCMS-QP2010 Series
Страница 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Страница 14: ...14 Do not increase the ion source or interface temperature above the level required for analysis...
Страница 27: ...27...
Страница 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Страница 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Страница 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Страница 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Страница 79: ...79 DC POWER A Block Diagram...
Страница 84: ...84 MAIN CTRL A Block Diagram...
Страница 85: ...85...
Страница 90: ...90 IS CTRL A Block Diagram...
Страница 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Страница 98: ...98 IG CTRL A PCB Block Diagram...
Страница 103: ...103 CP6 24V normal range 22 7 to 25 3VDC CP7 24V normal range 22 7 to 25 3VDC CP12 5V normal range 4 8 to 5 3VDC...
Страница 109: ...109 DI CTRL A block diagram...
Страница 145: ...121 5 Remove the IS CTRL A Board Unit Disconnect the two connectors Remove the three screws and the spacer screw...
Страница 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Страница 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Страница 154: ...130 10 Preparation of new TW300 New TW 300 Removed TW300 Check the name plate Old type New type...
Страница 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Страница 158: ...134 Hook the right side hooks on the neck of inlet Attach the left side screws and hooks Fasten lightly the 4 screws...
Страница 160: ...136 Remove the cardboard Completion of TMP mounting...
Страница 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Страница 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...
Страница 164: ...140 14 Place the covers Place the front cover Place the front door...