83
4. Checking the Pirani gauge measurement control circuits
Check the operational amplifier power supply voltage.
CP9 +5V
Confirm that circuit output voltage is in the correct range.
CP8 0.47 to 2.84V (1 to 300Pa) (Voltage decreases as the pressure increases.)
CP8 0.07V max. Discontinuity in Pirani gauge.
Check the Pirani gauge resistance.
Disconnect the Pirani gauge cable from the MAIN CTRL-A PCB connector J5 and measure
the resistance between Pins 1 and 2. The normal resistance is 10 to 15
Ω
.
5. Checking the ion-gauge vacuum measurement circuit
If measurements seem incorrect, check the output from the ion-gauge collector current
amplifier circuit.
CP6 [IG1]
A degree of vacuum in the range 1.3x10
-3
to1.3x10
-1
Pa is determined from the voltage V
CP6
.
Degree of vacuum (Pa) = 3.33x10
-2
x V
CP6
When the degree of vacuum exceeds 1.3x10
-1
Pa (V
CP6
≥
4V), the filament is turned off for
protection.
CP7 [IG100]
A degree of vacuum below 1.3x10
-3
Pa is determined from the voltage V
CP7
.
Degree of vacuum (Pa) = 3.28x10
-4
x V
CP7
A discontinuity in the ion-gauge measurement bulb or connection cable is evaluated if V
CP7
is 0.1V or less, and the ion-gauge filament is turned off. Inspect the cable between the
measurement bulb and connector J4 on the MAIN CTRL-A PCB.
Check the ion-gauge control circuit (IG CTRL-A) if the ion-gauge measurement bulb filament
does not light.
Содержание GCMS-QP2010 Series
Страница 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Страница 14: ...14 Do not increase the ion source or interface temperature above the level required for analysis...
Страница 27: ...27...
Страница 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Страница 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Страница 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Страница 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Страница 79: ...79 DC POWER A Block Diagram...
Страница 84: ...84 MAIN CTRL A Block Diagram...
Страница 85: ...85...
Страница 90: ...90 IS CTRL A Block Diagram...
Страница 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Страница 98: ...98 IG CTRL A PCB Block Diagram...
Страница 103: ...103 CP6 24V normal range 22 7 to 25 3VDC CP7 24V normal range 22 7 to 25 3VDC CP12 5V normal range 4 8 to 5 3VDC...
Страница 109: ...109 DI CTRL A block diagram...
Страница 145: ...121 5 Remove the IS CTRL A Board Unit Disconnect the two connectors Remove the three screws and the spacer screw...
Страница 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Страница 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Страница 154: ...130 10 Preparation of new TW300 New TW 300 Removed TW300 Check the name plate Old type New type...
Страница 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Страница 158: ...134 Hook the right side hooks on the neck of inlet Attach the left side screws and hooks Fasten lightly the 4 screws...
Страница 160: ...136 Remove the cardboard Completion of TMP mounting...
Страница 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Страница 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...
Страница 164: ...140 14 Place the covers Place the front cover Place the front door...