18 Error Messages
18.1 Error Messages
293
GC-2014
Abnormal component operation
Verify whether gas is supplied stably at the required pressure. If there is no problem in the
gas supply, the control system, such as the APC, may have failed. Turn off the system and
contact your Shimadzu representative.
18.1.4
Communication errors
External device communication errors (i.e., Chromatopac)
One of these messages appears during a link failure or communication failure. (When a
communication error occurs, the link is automatically disconnected.)
When one of these messages appears, check the connection status, and reset the link.
Code
Message
CS
2144
CAR1. L DAFC is out of control
Yes
2145
CAR1. R DAFC is out of control
Yes
2146
CAR2. L DAFC is out of control
Yes
2147
CAR2. R DAFC is out of control
Yes
2148
AMC.L is out of control
Yes
2149
AMC.L is out of control
Yes
2150
AMC.R is out of control
Yes
2151
AMC1 is out of control
Yes
2152
AMC2 is out of control
Yes
2153
AMC3 is out of control
Yes
2154
AMC4 is out of control
Yes
2155
AMC5 is out of control
Yes
2156
AMC6 is out of control
Yes
2157
AMC7 is out of control
Yes
2158
AMC8 is out of control
Yes
2159
AMC9 is out of control
Yes
2160
AMC10 is out of control
Yes
2161
CAR1 SAFC is out of control
Yes
2162
CAR2 SAFC is out of control
Yes
Code
Message
CS
4001
Time out
No
4002
Parity error
No
4003
Message is not accepted
No
4004
Data is invalid
No
4005
Command is invalid
No
4006
Data is out of range
No
4007
TRS port is shut down
No
4008
TRS file error
No
Содержание GC-2014
Страница 4: ...This page is intentionally left blank ...
Страница 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Страница 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Страница 42: ...2 Before Use 2 6 Connecting the Relay Terminals 30 GC 2014 This page is intentionally left blank ...
Страница 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Страница 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Страница 130: ...11 Creating an Oven Temperature Program 11 2 Temperature Programs 118 GC 2014 This page is intentionally left blank ...
Страница 164: ...12 Injection Port 12 5 Setting the Flow Rate Parameters 152 GC 2014 This page is intentionally left blank ...
Страница 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Страница 318: ...306 GC 2014 This page is intentionally left blank ...